background image

PSS Polymer Standards  
Service GmbH 

In der Dalheimer Wiese 5 
55120  Mainz  

|

 Germany

Polymer Standards  
Service-USA, Inc.

160 Old Farm Rd, Suite A 
Amherst | MA 01002 | USA

Phone

  +49 6131 96239-0

Fax

 

+49 6131 96239-11

E-Mail

 

[email protected]

Web

  

www.pss-polymer.com

Phone

  +1 413 835-0265

Fax

 

+1 413 835-0354

E-Mail

 

[email protected]

Web

  

www.pss-polymer.com

Tip: When using PSS MCX columns with laser light scattering detectors, it is recommended that 
they be purchased pre-equilibrated for light scattering (Lux version, precolumn P/N 299-2200, 
analytical P/N 299-2201). Extra steps must be undertaken to obtain optimum low noise baselines. 
Filtering solvents under vacuum through a 0.2µm filter and avoiding the use of salts and other 

additives in the eluent can improve performance

 

Noting the direction of flow etched on the column, connect the inlet fitting with the injector or preceding 
column using the appropriate nuts and new ferrules at each end of the tubing. Ensure that the capillary 
is as far as possible into the injector or column head before tighten the nut. Do not overtighten. The 
nut should be tightened enough so that the ferrule bites into the capillary tubing and no leak occurs. 
Excess force should not be required. 
 
To prevent trapped air being pumped into the column, unscrew the nut at the column inlet and slowly 
pump a few drops of eluent to waste before reconnecting the column. 

2.7 Connecting Columns in Series 

For connections between columns use the column connector delivered with each PSS MCX column. 
The precolumn should be connected to the injector and the remaining pore sizes connected in 
sequence of increasing pore size so that the largest porosity is closest to the detector. Each time a 
column is added, follow the procedure 2.6 above. 

2.8 Connecting the Column to the Detector 

Connect the outlet fitting of the column to a long piece of 1/16” tubing to take the eluent to the waste. 
Pump 3-4 column volumes of eluent through the column to waste at the recommended operational 
flow rate. Check for any leaks. Stop the flow, remove the waste tubing from the outlet and connect to 
the detector using the appropriate nuts and new ferrules at each end of the tubing. Apply the desired 
flow rate, again checking for leaks. As soon as the detector has stabilized, the column is ready for use.  

2.9 Testing Column Performance 

To verify correct installation, the column can tested by measuring the plate count/meter and the 
resolution to compared with the CoA noting that the measurements are a reflection of the dispersion in 
the whole system and not just the column alone. See the 

Full User’s Guide

 for details. Also note the 

back pressure in the chosen eluent. These parameters can be used to monitor ongoing column 
performance and to diagnose problems. 

3.  Usage Guidelines 

3.1 Eluent Preparation 

Use high quality HPLC grade eluents, free from large particles filtered using a 0.45µm filter if required. 
De-gas the eluent thoroughly, preferably using an online degasser.

 

 
3.2 Transferring to different solvents  

Only use PSS MCX columns with compatible eluents. 

Prepare the GPC/SEC system with the new eluent. Connect the column/column set to the injector and 
attach some 1/16” tubing to the outlet directed to waste. Pump approx. two column volumes of the 
eluent at a tenth the optimum flow rate through the column(s) and then slowly increase the flow rate to 
half the optimum flow rate and pump a further eight column volumes e.g. for 300mm x 8.0mm 
columns, start at 0.1mls/min and increase to 0.5mls/min. Stop flow and connect the column(s) to the 
detector. Start flow and increase the flow rate step by step up to the desired value. Ensure that all 
detector lines are completely purged and that the system is thoroughly equilibrated before making 
measurements. 

Отзывы: