![Proton HOGEN GC4800 Скачать руководство пользователя страница 7](http://html1.mh-extra.com/html/proton/hogen-gc4800/hogen-gc4800_installation-and-maintenance_1620406007.webp)
GC 4800 Desiccant Holder
Installation and Maintenance Manual
PD-0100-0063 Rev A
CONFIDENTIAL – Do not copy
Page 7 of 11
3.2
Desiccant Cartridge Installation
DO NOT DISSASSEMBLE THE HOLDER ASSEMBLY. DO NOT
LOOSEN THE NUTS SECURING THE TIE RODS. SERVICE
DESICCANT CARTRIDGE BY REMOVING THE APPROPRIATE
FASTENERS OUTLINED IN THIS SECTION.
1.
The service cap is located n the bottom of the desiccant holder. The cap is
secured in place with 6 screws, Figure 4. Using a ¼” hex drive. Remove the six
(6) screws.
Figure 4 Cartridge Holder Bottom with Serviceable Screws
2.
With the cap removed, prepare the desiccant cartridge, Part Number: 02-3001-
0040, for installation by removing the peel back lids on both ends of the cartridge,
Figure 5. Do not remove the ends from the cartridge until it is ready to be used
within the desiccant holder.
Remove Only
These Screws