Propaq CS Directions for Use
99
Ref
erence
Specifications
ECG Specifications
The ECG channel meets all the requirements for Cardiac Monitors Heart Rate Meters and Alarms
specified ANSI/AAMI EC13-1992, except for Standardizing Voltage (section 3.2.9.9). The channel also
meets the American National Standard, Safe Current Limits for Electromedical Apparatus
(ANSI/AAMI ES1-1993).
ECG Specifications
Characteristic
Specification
Connector
AAMI 6 pin or Hewlett-Packard compatible 12-pin style connector
(optional). See illustration on page 100.
Selectable Leads
I, II, III, aVR, aVL, aVF, V
Lead Fault Indicator
LA, LL, RA, RL, C, multiple
ECG Size (sensitivity) in mV/cm
4, 2, 1, 0.5, 0.2
Display Sweep Speeds
12.5, 25, and 50 mm/s
QRS Tone Volume
High, Low, Medium, Off
QRS Tone Frequency
900 Hz for Propaq CS monitor without Expansion Module, 665 Hertz
when equipped with SpO2 but SpO2 not being monitored; variable
pitch with SpO
2
option and SpO
2
being monitored
Bandwidth:
MONITOR
EXTENDED
Adult Mode:
0.5 to 40 Hz
Pediatric Mode:
0.5 to 120 Hz
Neonatal Mode:
0.5 to 120 Hz
Adult Mode:
0.05 to 40 Hz
Pediatric Mode:
0.05 to 120 Hz
Neonatal Mode:
0.05 to 120 Hz
(see Real-Time ECG Analog/Defib Sync specification)
Sample Rate
364 Hz
Input Protection
Electrosurgery and defibrillator protected when used with specified
ECG cables. All models also include electrosurgery interference
suppression.
Lead Fail Sense Current
50 nA dc for active leads
100-200 nA dc for driven lead, depending on number of electrodes
attached
Tall T-wave Rejection
Meets AAMI (USA) EC13-1992, section 3.1.2.1.c, for 1.2 mV T-wave
and 1 mV QRS using AAMI test waveform.
Common Mode Rejection
<1 mV p-p RTI for 10V rms, 50/60 Hz input, 200 pF source
impedance, input unbalanced, FILTER function OFF
<0.1 mV p-p RTI for 10V rms, 50/60 Hz input, 200 pF source
impedance, input unbalanced, FILTER function ON
Input Impedance
>2.5 M
Ω
differential @ 60 Hz
Input Range (ac)
10 mV peak to peak
Input Range (dc)
Up to ±300 mV
System Noise
≤
30 µV peak-to-peak, RTI, with all inputs = 47K in parallel with 0.047
µF.
QRS Detector
Adult or Pediatric Amplitude Range: 0.22 to 5.0 mV (RTI)
Neonatal Amplitude Range: 0.1 to 5.0 mV (RTI)
Neonatal and Pediatric Width Range (Duration):
40 to 120 ms
Adult Width Range (Duration): 70 to 120 ms
Heart Rate Range
25 to 350 beats per minute (measurement)
25 to 300 beats per minute (display)
Heart Rate Meter Response Time
Responds to change in heart rate within 5 to 9 seconds depending
on physiological waveform. (As measured per AAMI standard EC
13-1992 clause 4.1.2.1 (f), including 3.1.2.1 parts f. and g.
waveforms.) Includes 1 second readout update interval.
Содержание Propaq 242
Страница 6: ...6 Protocol Systems Inc...
Страница 18: ...18 Protocol Systems Inc...
Страница 26: ...26 Protocol Systems Inc...
Страница 50: ...50 Protocol Systems Inc...
Страница 52: ...52 Protocol Systems Inc...
Страница 64: ...64 Protocol Systems Inc...
Страница 66: ...66 Protocol Systems Inc...
Страница 72: ...72 Protocol Systems Inc...
Страница 74: ...74 Protocol Systems Inc...
Страница 78: ...78 Protocol Systems Inc...
Страница 84: ...84 Protocol Systems Inc...
Страница 86: ...86 Protocol Systems Inc...
Страница 94: ...94 Protocol Systems Inc...
Страница 120: ...120 Protocol Systems Inc...