INS8800 Rev. D
15
Protimeter MMS2 Instruction Manual
4.6 Logging - Selection and Use
The MMS2 supports both continuous and manual logging.
4.6a Manual Logging
If
is pressed at any of the measurement screens, the data and
Timestamp
at
that instant will be logged and a
RECORD SAVED
message will be displayed on
the bottom bar.
4.6b Continuous Logging
Continuous logging is used to sample and store data continuously. Continuous
logging is enabled either by setting logging parameters through the Keypad or
through a PC using
MMS2 Logging Software
. (see
Section 4.7g, “Set Logging
Parameters (Optional PC Logging Software Required)”, on page 19
.)
After the logging parameters are saved, logging starts after the
START AFTER
minutes have elapsed. The logging icon
is displayed on the top right
corner of the screen while logging is active.
When logging is in progress, an option to stop logging is provided under the
SETTINGS
menu. Logging can be stopped either by selecting
SELECT MODE ->
SETTINGS -> STOP LOGGING
in the instrument, by clicking
STOP LOGGING
in
the MMS2 Logging Software or when the instrument is turned
OFF
.
Содержание MMS2