V4.0_2021-05-26 

PROGNOST XS

 Instructions for use 

5034-0-0002 

 

 

 

13  of 25 

 

4

 

Handling 

4.1

 

 Requirements before and during Operation 

It must be ensured that the surfaces that come into contact with patients are disinfected before the X-
ray examination of each patient (see chapter 5.3). 
 

4.2

 

Operation

 

of

 

PROGNOST XS  

 

Exposures with the PROGNOST XS 

Following procedure must be observed to prevent collisions with the PROGNOST XS, the image 
receiver or other furniture.  

 

X-ray device, e.g. swivel arm system, to the 0 degree position 

 

Set the height of the X-ray image receptor to match the height of the table  

 

Move the table to the intended exposure position for inspection 

 

Move the table to a position in which the patient can climb onto  the surface as easily as possible  

 

Operate the brake wheel locks and check that the wheels are effectively braked 

WARNING! 

Before ascending or descending of the patient, the brake rollers must be 
locked. All rollers can be braked, but at least two rear rollers (furthest away 
from the ascending patient) must be locked.   

 

 

NOTE 

The PROGNOST XS is not suitable for patient transport.  
It is only used for positioning the patient during examination.  
The ascending and descending of the patient has only be done while fixed brakes 
and in the center of the table top, under supervision or assistance of the examiner as 
otherwise there is a risk of tipping over!  
Provide patients with assistance when ascending.  

 

 

Provide patients with assistance when ascending. In order to achieve the best possible support of 
the table, the patient should climb onto the table in the middle of the table top  

 

Position the patient for the exposure. If necessary (e.g. open wounds), cover the table surface with 
suitable cloths or disposal care pads 

 

WARNING! 

Danger of squeezing at the table edges and danger of trapping on and below 
the table top!  

When the table is moved horizontally, extremities can be squeezed between the 
edge of the table and a stationary obstacle (wall, column, X-ray unit). 
Therefore, when moving the PROGNOST XS, make absolutely sure that neither the 
patient nor the personnel are standing in the direction of the movement.   
Pay particular attention that none of the patients extremities protrude over the table 
top.  

 

 

Release the brake wheel locks and move the PROGNOST XS to the exposure position 

 

Lock the brakes again 

 

Perform exposures according to the operating instructions of the X-ray unit 

 

After the examination, release the fixing brakes and move the table to a position where the patient 
can easier get off the table  

 

Lock the brakes and assist the patient in getting off the table  

Содержание 7034-0-0000

Страница 1: ...y Telephone 49 0 7062 92 55 0 E Mail protec protec med com Version 4 0 Issued 2021 05 26 Subject to alterations PROGNOST XS Basic diagnostic X ray system table non powered Model ID 7034 0 0000 Basic U...

Страница 2: ...ithout the express written consent of PROTEC s law department Knowledge of violations of these regulations must be reported immediately to PROTEC GmbH Co KG 2021 PROTEC GmbH Co KG Oberstenfeld Comment...

Страница 3: ...onnel 11 Crushing and Collision Hazards 11 Explosion Protection 11 Interaction with External Devices 11 Electromagnetic Environment and the Interference of Devices 11 3 Control Elements and Device Dis...

Страница 4: ...7 1 Dimensions 21 7 2 Attenuation Equivalent 21 Protection Type and Protection Class 22 7 3 Environmental Conditions 22 Environmental conditions during operation 22 Environmental conditions for shippi...

Страница 5: ...f the equipment as of the date of manufacture Revisions to the equipment subsequent to the date of manufacture will be addressed in service updates distributed of the PROTEC GmbH Co KG Technical Servi...

Страница 6: ...ccompanying documents is required to carefully read through and carefully consider the instructions warnings and cautions contained therein before starting operation Even if you have already operated...

Страница 7: ...d Network System Requirements As a stand alone product the PROGNOST XS has no hardware or network connection and therefore no hardware or network requirements Installation NOTE The installation of the...

Страница 8: ...X ray has been given by a physician with the necessary expertise in radiation protection There are no general or fundamental restrictions on the patient group regarding age gender origin and patient c...

Страница 9: ...the European Parliament and of the Council of 5th April 2017 concerning medical devices including all applicable corrigenda The declaration of conformity is available upon request from PROTEC PROTEC G...

Страница 10: ...E After the initial installation the commissioning must be recorded with the PROTEC acceptance protocol FB 04 07A4 NOTE The commissioning of the PROGNOST XS can only be done if all safety notes and op...

Страница 11: ...ion Crushing and Collision Hazards WARNING Ensure that while operating the moving parts of the PROGNOST XS no persons or objects finds themselves in the obvious danger zone of the device Failure to do...

Страница 12: ...ice 3 2 Emergency stop switch of the PROGNOST XS The PROGNOST XS does not have an emergency stop switch as it is a non active medical device 3 3 Control elements and displays of the PROGNOST XS 1 Brak...

Страница 13: ...the patient has only be done while fixed brakes and in the center of the table top under supervision or assistance of the examiner as otherwise there is a risk of tipping over Provide patients with as...

Страница 14: ...05 26 PROGNOST XS Instructions for use 5034 0 0002 14 of 25 4 3 Function of the PROGNOST XS Switching the PROGNOST XS ON and OFF The PROGNOST XS does not have a main switch as it is a non active medic...

Страница 15: ...is usually done only with a dry cloth Do not use any corrosive dissolving or abrasive cleaning agents as this may damage the surfaces of the unit or the coating Clean device surfaces and painted parts...

Страница 16: ...PROTEC customer service or a service authorized by PROTEC to ensure the safe and reliable functioning of the device The maintenance intervals depend on the frequency of use In the event that the sche...

Страница 17: ...100 Patients day Every 4 years for systems with more than 100 patients day Visual control Functional control Exchange Exchange 1 Safety Inspection OK Not OK Note 1 1 Check the parts of the X ray syst...

Страница 18: ...S Instructions for use 5034 0 0002 18 of 25 Final Report No safety or function relevant problems detected Next maintenance in 6 12 ____ months Remarks notes Inspector Date Signature Doctor Doctors off...

Страница 19: ...customer service or an authorized service provider After the product has reached the end of its product life further use is at the operator s own risk Parts with direct patient contact Part Definition...

Страница 20: ...V4 0_2021 05 26 PROGNOST XS Instructions for use 5034 0 0002 20 of 25 6 Power supply NOTE The PROGNOST XS does not require power supply...

Страница 21: ...brakes of the wheels are mechanically operated 7 2 Attenuation Equivalent WARNING The PROGNOST XS attenuation factor must be considered during the acceptance test The aluminium attenuation equivalent...

Страница 22: ...contain any applied parts 7 3 Environmental Conditions Environmental conditions during operation Ambient Temperature 10 C to 40 C Relative humidity 30 to 70 non condensing Atmospheric pressure 700 hPa...

Страница 23: ...ions 8 1 Symbols Air pressure limit Temperature limit Humidity limit Keep dry Fragile Handle with care This way up Attention observe warnings and precautions Refer to the manual CE marking Manufacture...

Страница 24: ...rical and Electronic Equipment Protective earth Do not exceed the maximum indicated weight Do not exceed the maximum indicated weight Table movements for exposure Longitudinal movement of the table Tr...

Страница 25: ...ble crushing hazards to fingers or hands during movements of the table or X ray unit 8 4 Position of the Symbols and Labels 8 5 Abbreviations mm Millimetre cm Centimetre lb Pound kg Kilogram C Degree...

Отзывы: