Power frequency
(50/60 Hz)
magnetic field
EN 61000-4-8
3 A/m
EN 60601-1-2
Test level
Power frequency magnetic fields
should be at levels characteristic of
a
Typical location in a typical
commercial or hospital environment.
NOTE: U
T
is the alternating supply voltage prior to application of the test levels
Immunity
Test
EN 60601-1-2
Test level
Compliance
level
Electromagnetic Environment -
guidance
Radiated RF
EN 61000-4-3
3 V/m
80 MHz to 2.5 GHz
3 V/m
Portable and mobile RF
communications equipment
should be used no closer to any
part of the
Equipment,
including
cables, than the recommended
separation distance calculated from
the equation applicable to the
frequency of the transmitter
Recommended separation
distance
P
d
×
=
2
.
1
80 MHz to
800MHz
P
d
×
=
3
.
2
800 MHz to
2.5GHz
P
d
×
=
2
.
1
Where
P
is the maximum output
rating of the transmitter in watts
(W) according to the transmitter
manufacturer and
d
is the
recommended separation distance
in meters (m).
Field strengths for fixed RF
transmitter, as determined by an
electromagnetic site survey, should
be less than the compliance level in
each frequency range
Interference may occur in the
vicinity of equipment marked with
the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structure, objects and people.
a
Fields strengths from fixed transmitters, such as base stations for radio telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an
electromagnetic site survey should be considered. If the measured field strength outside the shielded
location in which the radiographic system is used exceeds [field strength] V/m, observe the
radiographic system to verify normal operation. If abnormal performance is observed, additional
measures may be necessary, such as relocating the radiographic system or using a shielded location
with a higher RF shielding effectiveness and filter attenuation