Movement and dimension
Table top dimensions (L x W):
226 cm x 75.5 cm or
200 cm x 75,5cm
Max. Patient weight (line load)
230 kg (standard)
250 kg (optional)
Table height:
697 mm
Tabletop transverse movement from center pos.:
±
150 mm
Tabletop longitudinal movement from center pos.:
±
330 mm (200 cm table top)
Tabletop longitudinal movement from center pos.:
±
470 mm (226 cm table top)
The brakes of the table top are actuated by a Bowden cable.
7.1.2
Total weight
The maximum total weight of the PROGNOST F without Bucky is 132 kg.
7.2
Attenuation Equivalent
WARNING!
The X-ray system table PROGNOST F can be delivered with different options
on the Grid Entity/Bucky.
The attenuation factor must be determined at the final inspection at the
customer. The variables like X-ray tube, Collimator etc. have influence to the
factor. The attenuation value of the components can be read out of the
accompanying documents of the component. The attenuation value has to be
determined at the technical specifications. If the limits can’t be kept please
inform PROTEC immediately. If additional accessories are use it has a negative
influence to the quality of the X-ray image.
The table top is defined as application part.
The aluminium attenuation equivalent of the table top is typically 0, 7 < 0, 8 Al mm for carbon / 0, 85
mm Al for composite fibre, according to EN 60601-1-3. Tested at 100 kV with a first half-value layer
thickness (HVLT) of 3, 7 mm Al and typically 0, 6 mm Al und <0,8mm Al according 21CFR § 1020-30 (n)
with 100 kV and a first half-value layer thickness (HVLT) of 2,7mm Al.
7.3
Protection Art and Protection Class
The PROGNOST F is consistent with a protection class 1 device and contains Applied Part of Type B
(according to EN 60601-1).
7.4
Environmental
7.4.1
Environmental conditions during operation
Ambient Temperature
+ 10°C to + 40°C
Relative humidity
30% to 75% (non-condensing)
Atmospheric pressure
700 hPa to 1060hPa
7.4.2
Environmental Conditions for Shipping and Storage
Ambient Temperature
- 10°C to + 70°C
Relative humidity
10% to 95% (non-condensing)
Atmospheric pressure
500 hPa to 1060hPa