29
MARK76X5002 Rev 3
Guidance and Manufacturer’s Declaration - Electromagnetic Immunity
The Luxura X5 Model 38X
is intended for use in the electromagnetic environment specified below.
The customer or the user of the Luxura X5 Model 38X should assure that it is used in such an
environment.
Immunity Test
IEC 60601 Test Level
Compliance Test
Electromagnetic
Environment - Guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
±2 kV, ±4 kV & ± 8 kV for
Contact Discharge
±2 kV, ±4 kV, ±8 kV and±
15kV for Air Discharge
±2 kV, ±4 kV & ± 8 kV for
Contact Discharge
±2 kV, ±4 kV, ±8 kV and±
15kV for Air Discharge
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrical fast transient/
burst IEC 61000-4-4
±2kV for power supply lines
±1kV for input/output lines
±2kV for power supply lines
±1kV for input/output lines
Mains power quality
should be that of a typical
commercial or hospital
environment.
Surge IEC 61000-4-5
AC Mains Line to Ground
±0.5kV, ±1kV and ±2kV
AC Mains Line to Line
±0.5kV and ±1kV
AC Mains Line to Ground
±0.5kV, ±1kV and ±2kV
AC Mains Line to Line
±0.5kV and ±1kV
Mains power quality
should be that of a typical
commercial or hospital
environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
<5% UT (100% dip in UT)
for 0.5 cycles
0% UT (100% dip in UT) for
1 cycle
70% UT (30% dip in UT) for
25/30 cycles
<5% UT (100% dip in UT)
for 250/300 cycles
<5% UT (100% dip in UT)
for 0.5 cycles
0% UT (100% dip in UT) for
1 cycle
70% UT (30% dip in UT) for
25/30 cycles
<5% UT (100% dip in UT)
for 250/300 cycles
Mains power quality
should be that of a typical
commercial or hospital
environment. If the user
of the Luxura X5 model
38X requires continued
operation during power
mains interruptions, it
is recommended that
the Luxura X5 model
38X be powered from an
uninterrupted power supply
or a battery.
Power frequency
(50/60Hz) magnetic
field
IEC 61000-4-8
30 A/m
30 A/m
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
NOTE: UT is the a.c mains voltage prior to application of the test level.
Содержание Luxura X5 38 SLi High Intensive
Страница 1: ...Luxura X5 38 SLi High Intensive 230V User Manual Model 38X 12 MINUTES MARK76X5002 Rev 3...
Страница 61: ...61 MARK76X5002 Rev 3 Emptying the A C Container Not all option may be available for your equipment...
Страница 66: ...66 MARK76X5002 Rev 3 1 Assembly...
Страница 67: ...67 MARK76X5002 Rev 3 2 3...
Страница 68: ...68 MARK76X5002 Rev 3 3 4...
Страница 69: ...69 MARK76X5002 Rev 3 6...
Страница 70: ...70 MARK76X5002 Rev 3 7...
Страница 71: ...71 MARK76X5002 Rev 3 8...
Страница 72: ...72 MARK76X5002 Rev 3 9...
Страница 73: ...73 MARK76X5002 Rev 3 10...
Страница 74: ...74 MARK76X5002 Rev 3 11...
Страница 75: ...75 MARK76X5002 Rev 3 12...
Страница 76: ...76 MARK76X5002 Rev 3 13...
Страница 77: ...77 MARK76X5002 Rev 3 14...
Страница 78: ...78 MARK76X5002 Rev 3 15...
Страница 79: ...79 MARK76X5002 Rev 3 16...
Страница 80: ...80 MARK76X5002 Rev 3 17 18...
Страница 81: ...81 MARK76X5002 Rev 3 19...
Страница 82: ...82 MARK76X5002 Rev 3 20...
Страница 83: ...83 MARK76X5002 Rev 3 21...
Страница 84: ...84 MARK76X5002 Rev 3 22 23...
Страница 85: ...85 MARK76X5002 Rev 3 24...
Страница 86: ...86 MARK76X5002 Rev 3 25 26...
Страница 87: ...87 MARK76X5002 Rev 3 27 28...
Страница 88: ...88 MARK76X5002 Rev 3 29...
Страница 89: ...89 MARK76X5002 Rev 3 30...
Страница 90: ...90 MARK76X5002 Rev 3 32 31...
Страница 91: ...91 MARK76X5002 Rev 3 33 34...
Страница 92: ...92 MARK76X5002 Rev 3 35...
Страница 93: ...93 MARK76X5002 Rev 3 36...
Страница 94: ...94 MARK76X5002 Rev 3 37 38...
Страница 95: ...95 MARK76X5002 Rev 3 39 40...
Страница 96: ...96 MARK76X5002 Rev 3 41 42...
Страница 97: ...97 MARK76X5002 Rev 3 43...
Страница 98: ...98 MARK76X5002 Rev 3 44 45...
Страница 99: ...99 MARK76X5002 Rev 3 46 47...