b. Place a strip of aluminum foil approximately 1 inch wide by 14 inches long in the test shoe.
Make sure that the foil strip runs the full length of the shoe from toe to heel.
(1)
Have the Test Subject don the shoe
(2)
Approximately 2 inches of foil should extend from the heel of the
shoe.
c. Place an alligator clip on the second test lead and a ach it to the foil strip protruding from
the right shoe.
d. While the Human Test Subject is wearing the test shoes, the subject stands on the hinged
test plate and fi rmly holds the Hand Wand in one hand.
e. Press
TEST
bu on on the meter, and read the resistance from the display.
f. Repeat at 100V if desired or necessary.
3. A er the Right Shoe resistance is recorded, repeat the en re procedure with the snap on the
le side of the hinged test plate and the second test lead clipped to the aluminum foil extend-
ing from the Le shoe.
A er the Le Shoe resistance is recorded, turn the meter
OFF
.
7