DEVICE OVERVIEW

Therapy Unit

The 

ProMOTION EV1 Therapy Unit features a multifunction interactive keypad 

and LCD display. The keypad can be used to power the unit on, select a program 
mode, start or stop a program and to power the unit down.

The LCD display indicates; power status, program selection, program status and 
error messages.

The Therapy Unit also contains a Pump Connection Port located adjacent to the 
LCD display. This is used to connect to the Pump unit.

DEVICE OVERVIEW

P10

Power ON / OFF 
/ Select menu

Go back

Next menu (Down)

Next menu (Up)

Pump  
connection port 
(Not for charging)

LCD display

Interactive keypad

P11

ProMOTION EV1

ProMOTION EV1

User Manual

User Manual

Содержание EV1

Страница 1: ...www promotion fitness USER MANUAL Version 1 0 ...

Страница 2: ...taking sports and general exercise The ProMOTION EV1 integrates cooling heating and compression therapy in a unique fully portable system Compression and temperature are easily programmed and adjusted at the touch of a button for specific prorgrams and user needs with no requirement for water ice or heavy equipment Please feel free to let us know what you think of our products we d love to hear fr...

Страница 3: ...301 489 17 V3 2 4 Radio Emissions This Therapy Unit has been approved against the following standards for Radio Emissions ETSI EN 300 328 V2 2 2 EN 62311 2008 Safety Testing This Therapy Unit has been approved against the following standards for Safety testing EN 60335 1 2012 A11 2014 EN 60335 2 32 2003 A2 2015 The Therapy Unit s battery pack has been approved against the following standards for S...

Страница 4: ...nuria cryoglobulinemia sickle cell disease serum cold agglutinins Caution Compression therapy with the ProMOTION EV1 device should be used only under the supervision of a licensed healthcare practitioner on patients Who have a wound in the affected region the wound must be dressed prior to use of the ProMOTION EV1 device Who have an acute unstable untreated fracture in the affected region CONTRAIN...

Страница 5: ...he device on an aircraft Handle the ProMOTION EV1 device with care Never operate the device with damaged or modified components or if the unit is otherwise not fully operational If you suspect battery leakage stop usage immediately Batteries should be disposed of carefully in line with local authority regulations Do not use any cables charging adapters or accessories other than those provided by P...

Страница 6: ... down The LCD display indicates power status program selection program status and error messages The Therapy Unit also contains a Pump Connection Port located adjacent to the LCD display This is used to connect to the Pump unit DEVICE OVERVIEW P10 Power ON OFF Select menu Go back Next menu Down Next menu Up Pump connection port Not for charging LCD display Interactive keypad P11 ProMOTION EV1 ProM...

Страница 7: ...tery The ProMOTION EV1 battery features a battery level indicator and button charging port and eject tab Depressing the indicator button will result in the level of charge being indicated 1 LED 25 charge 2 LED 50 charge 3 LED 75 charge 4 LED 100 charge USB connection Bayonet fitting Battery level indicator Alignment slot Charging port Battery level button P12 P13 ProMOTION EV1 ProMOTION EV1 User M...

Страница 8: ...ION EV1 system Small Medium S M and Medium Large M L Therapy Unit cradle Pump Unit port Handle Adjustment array USING YOUR PROMOTION EV1 DEVICE Battery charging and usage The battery should be fully charged before use to ensure maximum performance Follow the procedure below before using the device 1 If the battery is installed in the Therapy Unit lift the eject latch on the battery The battery wil...

Страница 9: ...f charge is indicated by how many LEDs pulse 1 LED 25 charge 2 LED 50 charge 3 LED 75 charge 4 LED 100 charge Charging is complete when the battery level indicator turns off 4 Unplug the charging cable from the battery charging port once the battery is charged Precautions Only use supplied charging cable The battery should never be covered during the charging cycle It is recommended that battery c...

Страница 10: ...dle Allow the spring to push back locking the Therapy Unit in place Precautions Check the Therapy Unit is secure by pulling it up and away from the cradle Do not obstruct the air inlet or exhaust at either end of the Therapy Unit USING YOUR PROMOTION EV1 DEVICE Installing Pump Unit The Pump Unit attaches to the Compression wrap via a simple bayonet fitting To attach it 1 Align the Pump Unit with t...

Страница 11: ...Therapy Unit over the injury site 2 Wrap the Compression wrap comfortably around the limb It should be secure enough not to slip but not tight 3 Locate the wrap handle into the adjustment array securing the wrap assembly in place 4 Push the Pump unit cable and clip into the USB port on the Therapy Unit until it clicks into place Precautions Check that the Pump Unit fits the bayonet port correctly ...

Страница 12: ...igate the menus Pressing the Up Down buttons will cycle through the menus or adjust the levels Use the Select and Go back buttons to make a selection or return to the previous menu Temperature compression and duration can be adjusted and selected USING YOUR PROMOTION EV1 DEVICE Fitting the universal strap 1 Align the Therapy Unit with the cradle and push it into place 2 If not already located push...

Страница 13: ... Then press the button again to stop the program NEED SCREENSHOTS HERE 4 The Compression wrap will start to deflate before the end of the program Once the program is complete the device will revert to standby mode and the wrap can be removed Once the wrap has deflated gently pull the handle free of the sizing array P24 P25 ProMOTION EV1 ProMOTION EV1 User Manual User Manual 1 Press and hold the Se...

Страница 14: ...u to create and repeat new therapy programs and access pre designed routines Use the keypad to navigate to each menu Pressing the up or down keys will adjust the setting Once you are happy with your choice press Select the device will retain your input After powering up the Therapy unit will display a series of menus enabling you to operate and program the device New History and Programs give acce...

Страница 15: ...m this screen in displayed Press and hold Go back to unlock the screen Press Go back again to stop the program Locked The device is under control of the app P29 ProMOTION EV1 ProMOTION EV1 User Manual User Manual MENU OVERVIEW New Selecting New will create a new program where temperature program duration and compression levels can be amended to suit 1 Use the keypad to navigate to each menu Temper...

Страница 16: ...m each other while the device is in use Doing so may cause inadvertant system error If the Pump connection cable or Pump Unit are disconnected at any stage simply reconnect it The Pump Unit will take a few moments to recalibrate before continuing the program The Therapy Unit will not recall incomplete or interrupted programmes If the battery is removed during a program the device will reset from s...

Страница 17: ...gram to select a predefined program Select New Program to define your own program Select Program History to access previously completed programs P32 P33 ProMOTION EV1 ProMOTION EV1 User Manual User Manual PROMOTION EV1 APP Default Programs The ProMOTION EV1 device and app allocates ten program slots Three of these are default programs created by ProMOTION EV1 the rest are available for users to cr...

Страница 18: ... Manual ProMOTION EV1 User Manual CLEANING MAINTENANCE STORAGE Therapy Unit Pump Unit The exterior of the device can be cleaned with a soft damp cloth Do not use any solvent based cleaners or abrasive materials Caution The ProMOTION EV1 is not a waterproof device Do not apply a direct stream of any liquid onto the device submerge it or allow any liquid to pool on the device Compression Wrap Strap ...

Страница 19: ... ProMOTION Customer Services Battery charge may be insufficient please recharge Ambient Temperature Humidity may be outside recommended range wait until within range Battery may be damaged or expired please contact ProMOTION Customer services Therapy Unit will not turn on or function as desired Therapy Unit will not achieve desired temperature Pump Unit does not operate as desired Compression Wrap...

Страница 20: ...P38 Email customerservices swellaway com www promotion fitness resources NOTES ...

Страница 21: ...P26 www promotion fitness ...

Отзывы: