III.D.
Operational Mode Setup (continued)
The Maxwell
®
16 Instrument has multiple operational modes, depending on the purification
procedure and the Maxwell
®
16 Purification Kit being used. Table 1 lists the hardware configuration
and operational mode requirements for the various Maxwell
®
16 Purification Kits. Please refer to the
Technical Bulletin or Technical Manual supplied with your Maxwell
®
16 Purification Kit for further
information on the operational mode setting required for your selected system.
Table 1. Hardware Configuration and Operational Mode Setup Requirements.
Please ensure that you select the operational mode that is required for your hardware configuration.
Failure to use the correct operational mode and hardware configuration will cause damage to the
instrument.
!
Promega Corporation ·
2800 Woods Hollow Road
·
Madison, WI 53711-5399 USA
·
Toll Free in USA 800-356-9526
·
Phone 608-274-4330
·
Fax 608-277-2516
· www.promega.com
Printed in USA.
Part# TM274
Revised 6/08
Page 17
Maxwell
®
16 Kit
Purification
Procedure
Firmware
Operational Mode
Hardware
Configuration
Maxwell
®
16 gDNA
Purification Kits
(Cat.# AS1010, AS1020,
AS1030)
gDNA
Research
(Rsch)
Standard Elution
Volume
(SEV)
Maxwell
®
16 Polyhistidine
Protein Purification Kit
(Cat.# AS1060)
Polyhistidine-
Tagged Protein
Research
(Rsch)
Standard Elution
Volume
(SEV)
Maxwell
®
16 Total RNA
Purification Kit
(Cat.# AS1050)
Total RNA
Research
(Rsch)
Standard Elution
Volume
(SEV)
Maxwell
®
16 LEV Total
RNA Purification Kits
(Cat.# AS1220, AS1225)
Total RNA
Research
(Rsch)
Low Elution
Volume
(LEV)
DNA IQ™ Reference
Sample Kit for Maxwell
®
16
(Cat.# AS1040)
Forensic gDNA
Reference Samples
Forensic
(Fnsc)
Standard Elution
Volume
(SEV)
DNA IQ™ Casework
Sample Kit for Maxwell
®
16
LEV (Cat.# AS1210)
Forensic gDNA
Casework Samples
Forensic
(Fnsc)
Low Elution
Volume
(LEV)