11
Promega Maxwell
®
CSC 48 Instrument | TM623 | Revised 6/20
1.10 Environmental Requirements (Operating, Shipping and
Storage Conditions)
Power Requirements:
100–240VAC, 50/60Hz, 4A
Temperature:
4–50°C (shipping/storage), 15–25°C (operation)
Humidity:
Up to 80% relative humidity, non-condensing
Operating Altitude:
<2,000 meters
The Maxwell
®
CSC 48 Instrument is intended for indoor use only. Wipe up spills immediately.
To avoid shortening the expected lifespan of the instrument, install in a location that meets the
following criteria:
• Locate on a sturdy, level surface.
• Avoid dusty areas.
• Choose a location that has good air circulation and is not exposed to direct sunlight.
• Avoid electrically noisy power sources (e.g., power generators).
• Do not install in a location where there is large temperature variability or high humidity.
• Do not position the instrument so that it is difficult to unplug from the power source.
• Do not place next to heat sources.
• Do not use near flammable gases or liquids.
• Do not place near other electrically sensitive instruments.
• Connect the Maxwell
®
CSC 48 Instrument and Tablet PC to an uninterrupted power supply
(UPS). This is to ensure continued operation during brief power outages that otherwise might
result in aborted instrument runs and lost samples.
Introduction
Содержание AS8000
Страница 66: ...65 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes Operating the Maxwell CSC 48 Instrument ...
Страница 70: ...69 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes Cleaning and Maintenance ...
Страница 74: ...73 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes ...
Страница 86: ...85 Promega Maxwell CSC 48 Instrument TM623 Revised 6 20 Notes ...