USER’S MANUAL. PROPOWER-5
02/2007
Page
17
En
g
lis
h
Figure 10.-
SCAN mode,
Ref. 20 dBm.
Figure 12.-
SCAN mode, power
detection function.
To access to this operating mode,
press the
SCAN
key [11]. The top line of the
screen will display the channel name (UL4),
the tuned frequency (10.5 MHz) and the level (-10
dBm) of the channel the marker is over. The marker
can be moved by turning the rotary selector [9].
The channel is selected when it is shaded.
To change the
span
(shown
bandwidth), press the rotary selector [9]
(the span field is activated) and then turn
it. The span is variable from the entire band (
full span
)
to 10 MHz in the following steps:
1, 5, 15, 30 MHz and
full span
.
By pressing the rotary selector [9]
again, you can change the reference
level, in other words, on turning the rotary
selecter [9] the graph is displaced vertically.
To modify the tuned channel, press the rotary
selector [9] again, the channel field will be activited and
then turning the rotary selector you will be able to
change it.
The attached figure shows an horizontal line at
-15 dBm. This line allows to establish judgements of
channel level acceptance easily. To activate and
define this
reference line
see the
Scan
configuration menu
(section 4.2.2.1).
In the case of having activated the broadband
power detection function (
POWER MEAS ->
∑
PWR
)
in the configuration menu of
SCAN
mode (See
section 4.2.2.1), on the display of
PROPOWER-5
will
appear the total power level on the instrument input.
This parameter indicates the total power that is
being transmitted by the connected cable to the
equipment in the band from 1 to 50 MHz.
The
SCAN
operating mode also allows channels to be programmed for use as
pilots in
TILT
operating mode, in the forward band (see section
4.2.5 TILT operating
mode
).
IMPORTANT
In order to achieve the maximum accuracy, the digital PLC channels power must
be measured always from the CHANNEL-FREQUENCY operating mode.
Figure 8.-
SCAN mode,
full span
.
Figure 9.-
SCAN mode,
span 10 MHz
.
Figure 11.-
SCAN mode with
reference line at -15 dBm.