The number of traverses required depends on the sensor type, the nominal diameter and the thickness of the 
pipe wall. We recommend the following types of mounting:

Two-channel operation

The transmitter is able to operate two independent measuring channels (measuring channel 1 and measuring 
channel 2). A pair of sensors is connected per measuring channel. Both measuring channels operate 
independently of one another and are supported by the transmitter to an equal extent. 

Two-channel operation can be used for the following measurements:
• Two-channel measurement = flow measurement at two separate measuring points
• Two-path measurement = redundant flow measurement at one measuring point

Two-channel measurement

The flow is measured at two separate measuring points in the case of two-channel measurement.

The measured values of the two measuring channels can be processed and displayed differently.
The following measured values can be output for two-channel measurement:
• Individual measured values per measuring channel (output independently of one another)
• The difference between the two measured values
• The sum of the two measured values

The two measuring channels can be configured individually. This makes it possible to independently configure 
and select the display, outputs, sensor type and type of installation.

Sensor Type

Nominal Diameter

Sensor Frequency

Sensor ID

Type of Mounting

 1)

Prosonic Flow W

DN 15 to 65 (½ to 2½")

6 MHz 

W-CL-6F*

2 traverses 

5)

DN 50 to 65 (2 to 2½")

2 MHz 

W-CL-2F*

2 (or 1) traverses 

2)

DN 80 (3")

2 MHz

W-CL-2F*

2 traverses

DN 100 to 300 (4 to 12")

2 MHz (or 1 MHz)

W-CL-2F*
W-CL-1F*

2 traverses 

3)

DN 300 to 600 (12 to 24")

1 MHz (or 2 MHz)

W-CL-1F*
W-CL-2F*

2 traverses 

3)

DN 650 to 4000 (26 to 160")

1 MHz (or 0.5 MHz)

W-CL-1F*
W-CL-05F*

1 traverse 

3)

1)

 The installation of clamp-on sensors is principally recommended in the 2 traverse type installation. This type of 
installation allows the easiest and most comfortable type of mounting and means that a system can also be mounted 
even if the pipe can only be accessed from one side. However, in certain applications a 1 traverse installation may be 
preferred. These include:

• Certain plastic pipes with wall thickness > 4 mm (0.16")
• Pipes made of composite materials such as GRP
• Lined pipes
• Applications with fluids with high acoustic damping

2)

If the pipe nominal diameter is small (DN 65 / 2½" and smaller), the sensor spacing with Prosonic Flow W can be too 
small for two traverse installation. In this case, the 1 traverse type of installation must be used.

3)

0.5 MHz sensors are also recommended for applications with composite material pipes such as GRP and may be 
recommended for certain lined pipes, pipes with wall thickness > 10 mm (0.4"), or applications with media with high 
acoustic damping. In addition, for these applications we principally recommend mounting the W sensors in 
a 1 traverse configuration.

4)

Insertion W sensors are mounted in a 1 traverse configuration 

ä

23.

5)

6 MHz sensors for appliations where flow velocity < 10 m/s 832.8 ft/s)

Содержание Prosonic Flow 93W

Страница 1: ...ly suitable for retrofit measurement flow monitoring improving measuring points Interface for easy integration into all common distributed control systems HART PROFIBUS DP PA FOUNDATION Fieldbus Featu...

Страница 2: ...bility 14 Operating conditions installation 14 Installation instructions 14 Inlet and outlet runs 15 Operating conditions environment 16 Ambient temperature range 16 Storage temperature 16 Degree of p...

Страница 3: ...oss sectional area The measuring system calculates the volume flow of the fluid from the measured transit time difference and the pipe cross sectional area In addition to measuring the transit time di...

Страница 4: ...n the table and you do not have the corresponding sound velocities for these fluids materials you can use the DDU18 and DDU19 sensors to determine the values Prosonic Flow 93 wall mount housing For mo...

Страница 5: ...sensors are located on the same side of the pipe A0001108 Sensor mounting arrangement A Mounting arrangement for measurement via one traverse B Mounting arrangement for measurement via two traverses D...

Страница 6: ...Type of Mounting 1 Prosonic Flow W DN 15 to 65 to 2 6 MHz W CL 6F 2 traverses 5 DN 50 to 65 2 to 2 2 MHz W CL 2F 2 or 1 traverses 2 DN 80 3 2 MHz W CL 2F 2 traverses DN 100 to 300 4 to 12 2 MHz or 1...

Страница 7: ...e Averaging function generally provides you with a more stable measured value The function is thus suitable for measurements under conditions that are not ideal e g short inlet runs The two measuring...

Страница 8: ...ctor 30 V DC 250 mA Time constant selectable 0 05 to 100 s Frequency output End frequency 2 to 10000 Hz fmax 12500 Hz On off ratio 1 1 pulse width max 10 s Pulse output Pulse value and pulse polarity...

Страница 9: ...A Error current FDE fault disconnection electronic 0 mA Signal encoding Manchester II Function blocks 8 Analog Input AI 1 Discrete Output 1 PID Output data volume flow channel 1 or channel 2 sound vel...

Страница 10: ...o 26 DP B PA FF with polarity protection Terminal No 27 DP A PA FF with polarity protection e Ground terminal for signal cable shield Fieldbus cable f Service adapter for connecting service interface...

Страница 11: ...t output HART 93 B Relay output 2 Relay output 1 Frequency output Current output HART 93 H PROFIBUS PA 93 J PROFIBUS DP 93 K FOUNDATION Fieldbus Flexible communication boards 93 C Relay output 2 Relay...

Страница 12: ...4 mm 2 0 16 Cable gland M20 1 5 Thread for cable entry NPT G A0008152 Cable gland for two connecting cables per cable entry Connecting cable sensor transmitter Only use the connecting cables supplied...

Страница 13: ...or at the measuring point 0 5 o r 3 mm s 1 5 o r 2 o r 3 mm s Measured error at the measuring point The measured error at the measuring point is made up of the measured error of the device 0 5 o r and...

Страница 14: ...l the sensors in a riser Note Entrained air or gas bubbles in the measuring tube can result in an increase in measuring errors For this reason avoid the following mounting locations Highest point of a...

Страница 15: ...installation range max 120 Inlet and outlet runs If possible install the sensor well clear of fittings such as valves T pieces elbows etc Compliance with the following inlet and outlet runs is require...

Страница 16: ...te It is permitted to insulate the sensors mounted on the pipe Mount the transmitter in a shady location and avoid direct sunlight particularly in warm climatic regions Storage temperature The storage...

Страница 17: ...ement Standard 40 to 80 C 40 to 176 F Optional 0 to 170 C 32 to 338 F DDU19 sensor accessories wall thickness measurement 20 to 60 C 4 to 140 F Medium pressure range nominal pressure No pressure limit...

Страница 18: ...0 81 53 L M N O P Q R S T 1 95 53 102 81 5 11 5 192 8 M5 20 2 6 5 1 Securing screw for wall mounting M6 screw head max 10 5 mm All dimensions in mm A B C D E F G H J K 8 46 9 84 3 56 6 27 5 31 3 54 1...

Страница 19: ...aid mounting mounting kits are available as accessories 28 A0001131 Pipe mounting Note To aid mounting mounting kits are available as accessories 28 A0001132 245 9 65 110 4 33 210 8 27 0 5 0 019 0 5...

Страница 20: ...Hauser Prosonic Flow W sensor DN 15 to 65 to 2 A0011502 Dimensions in SI units Dimensions in US units B A C E F D A B C D E F 72 331 39 28 233 450 All dimensions in mm A B C D E F 2 83 13 03 1 541 1...

Страница 21: ...0 5 H I Depends on the measuring point conditions pipe fluid etc Dimension H can be determined Via transmitter programming Quick Setup or FieldCare Online Applicator Pipe outer diameter All dimension...

Страница 22: ...ring point conditions pipe fluid etc Dimension G can be determined Via transmitter programming Quick Setup or FieldCare Online Applicator Pipe outer diameter All dimensions in mm A B C D E F 2 20 2 44...

Страница 23: ...suring point conditions pipe fluid etc Dimensions F and G can be determined Via transmitter programming Quick Setup or FieldCare Online Applicator All dimensions in mm A B C D E F G 2 28 5 91 2 56 25...

Страница 24: ...e fluid etc Dimensions F G and H can be determined Via transmitter programming Quick Setup or FieldCare Online Applicator All dimensions in mm A B C D E F G H 2 28 5 91 3 15 25 Pipe outer diameter Sen...

Страница 25: ...cket textile or stainless steel 1 4301 304 Sensor contact surfaces chemically stable plastic Prosonic Flow W insertion version Sensor holder stainless steel 1 4301 Sensor housing stainless steel 1 430...

Страница 26: ...k Setup menus for straightforward commissioning Language group Language groups available for operation in different countries Western Europe and America WEA English German Spanish Italian French Dutch...

Страница 27: ...ATION Fieldbus H1 Interoperability Test Kit ITK revision status 5 01 device certification number on request The device can also be operated with certified devices of other manufacturers Physical Layer...

Страница 28: ...ppropriate plug in point modules for converting the current input output configuration to a new version DK9UI Sensor W DN 15 to 4000 to 160 Clamp on version DN 15 to 65 to 2 20 to 80 C 4 to 176 F 6 0...

Страница 29: ...nic Flow W DN 15 to 65 to 2 Conduit adapter incl cable entry M20 1 5 Conduit adapter incl cable entry NPT Conduit adapter incl cable entry G Prosonic Flow W DN 50 to 4000 2 to 160 Conduit adapter incl...

Страница 30: ...browser and or WAP cellular phone Limit value monitoring with alarm signaling via e mail or SMS Synchronized time stamping of all measured values Remote diagnosis and remote configuration of connected...

Страница 31: ...ieldbus BA078D and BA079D Supplementary documentation on Ex ratings ATEX FM CSA IECEx NEPSI Registered trademarks HART Registered trademark of HART Communication Foundation Austin USA PROFIBUS Registe...

Страница 32: ...ational Endress Hauser Instruments International AG Kaegenstrasse 2 4153 Reinach Switzerland Tel 41 61 715 81 00 Fax 41 61 715 25 00 www endress com info ii endress com TI00084D 06 EN 13 11 71136723 F...

Отзывы: