3.3  CURRENT MEASUREMENT   

3.3.1 

Insert the black test lead into “COM” jack, the red one to “mA” or 
“10A”jack.  

3.3.2  Set  the  range  knob  to  a  proper  DC  or  AC  mA/A  range,  If  the 

measured voltage is unsure beforehand, set the range knob to the 
highest  range,  then  reduce  it  gradually  until  get  the  highest 
resolution readings.  

3.3.3  Connect the test leads to the circuit under test, the LCD displays 

the measured voltage value.  

NOTE:             
a. 

If LCD displays “OL”, it means over range, set the range knob to a 
higher range.  

b. 

When measuring current,

 

mA hole should not  exceed 200mA,10A 

hole should not exceed 10A(test time should be less than 10 sec.)  

  

3.4  RESISTANCE MEASUREMENT  

3.4.1 

Insert the black test lead to “COM” jack and the red one to

 

“V/Ω/Hz” 

jack.  

3.4.2  Set the range knob to a proper resistance range, connect the test 

leads across to the resistance being measured.  

NOTE:  
a. 

If  the  resistance  value  being  measured  exceeds  the  max  value  of 
the  range  selected,  LCD  displays  "OL",  ,  set  the  range  knob  to  a 
higher range. When the resistance is over 1M

Ω, the meter may take 

a  few  seconds  to  stabilize.  This  is  normal  for  high  resistance 
readings.   

b. 

When input terminal is in open circuit, overload displays.  

c. 

When measuring in-line resistance, be sure that power is off and all 
capacitors are discharged completely.  

d. 

Do not input any voltage at this range.  

  

3.5  CAPACITANCE MEASUREMENT  

3.5.1 

.Insert the red test lead to “V/Ω/Hz”

 

terminal and the black one to 

“COM” jack.  

3.5.2  Set the range knob to a proper capacitance range, connect the test 

leads  to  the  capacitor  being

 

measured

note:  the  polarity  of  red 

test lead is “+”

.  

NOTE:   
a. 

If  the  resistance  value  being  measured  exceeds  the  max  value  of 
the range selected, LCD displays "OL".  

Содержание MT-1705

Страница 1: ...MT 1705 3 1 2 True RMS Multimeter User s Manual 1st Edition 2014 Copyright by Prokit s Industries Co Ltd...

Страница 2: ...t leads are connected correctly whether the insulation is good when measuring over 36DCV or 25ACV 3 Remove the test leads when changing function and range 4 To select correct function and range make s...

Страница 3: ...200mV 0 5 3 100uV 2V 1mV 20V 10mV 200V 100mV 600V 1 0 10 1V Input resistance All ranges 10 M Overload protection 250V DV or AC peak value at 200mV range 600V DC or AC peak value at other ranges 2 2 2...

Страница 4: ...2 2 4 ACA True RMS measurement Range Accuracy Resolution 40Hz 200Hz 200Hz 1kHz 200mA 1 5 15 1 5 15 sine and triangular wave 8 0 15 other wave 100uA 10A 3 0 10 3 0 10 sine and triangular wave 10mA Max...

Страница 5: ...RESOLUTION 20nF 3 5 20 10pF 200nF 100pF 2uF 1nF 20uF 5 0 10 10nF 200uF 100nF 2000uF 1uF Overload protection 250V DC or AC peak value NOTE when over range it can keep on measuring OL will not show on L...

Страница 6: ...n and range 3 1 5 GND 3 1 6 10A current test jack 3 1 7 pole jack of voltage resistance diode capacitance and resistance 3 2 VOLTAGE MEASUREMENT 3 2 1 Insert the black test lead to COM jack the red on...

Страница 7: ...range knob to a proper resistance range connect the test leads across to the resistance being measured NOTE a If the resistance value being measured exceeds the max value of the range selected LCD dis...

Страница 8: ...less a reading is possible but maybe over range b Shielded cable is recommended when measuring small signals under noisy conditions c Be careful when measuring high volt circuit d Do not input a volta...

Страница 9: ...meter and should only be serviced by factory personnel 4 1 Do not operate or store the instrument in high temperature or high humidity place and do not work closed to flammability substance or explosi...

Страница 10: ...9 NO CURRENT INPUT Replace fuse BIG ERROR Replace battery PROKIT S INDUSTRIES CO LTD http www prokits com tw E mail pk mail prokits com tw 2014 Prokit s Industries Co LTD All rights reserved 2014001 C...