3302F Mainframe Operation Manual 49
OCP: STEP
Command Syntax
:
[PRESet
:
] OCP:STEP {SP}{NR2}{
;
NL}
[PRESet
:
] OCP:STEP
?
{
;
NL}
Purpose:
Set and read the increasing value of OCP Test
Description:
This command is used for setting the increasing value (I-STEP) of OCP
Test.
OCP: STOP
Command Syntax
:
[PRESet
:
] OCP:STOP {SP}{NR2}{
;
NL}
[PRESet
:
] OCP: STOP? {
;
NL}
Purpose:
Set and read the maximum value of OCP Test
Description:
This command is used for setting the maximum value (I-STOP) of OCP
Test.
VTH
Command Syntax
:
[PRESet
:
] VTH {SP}{NR2}{
;
NL}
[PRESet
:
] VTH ? {
;
NL}
Purpose:
Set and read the value of the Threshold Voltage
Description:
This command is used for setting a minimum Threshold Voltage for the
OCP/OPP tests. If the measured voltage is below the threshold voltage
And the OCP/OPP test started the test will not run and an error will be
Flagged.
OPP: START
Command Syntax
:
[PRESet
:
] OPP:START {SP}{NR2}{
;
NL}
[PRESet
:
] OPP:START
?
{
;
NL}
Purpose:
Set and read the initial value of OPP Test
Description:
This command is used for setting the initial value (P-START) of the OPP
Tes
t.
OPP: STEP
Command Syntax
:
[PRESet
:
] OPP:STEP {SP}{NR2}{
;
NL}
[PRESet
:
] OPP:STEP
?
{
;
NL}
Purpose:
Set and read the increasing value of OPP Test
Description:
This command is used for setting the power increments which the
OPP test will follow between the P-START and P-STOP values.
OPP: STOP
Command Syntax
:
[PRESet
:
] OPP:STOP {SP}{NR2}{
;
NL}
[PRESet
:
] OPP:STOP
?
{
;
NL}
Purpose:
Set and read the maximum value of OPP Test
Description:
This command is used to set the maximum power value (P-STOP) of
The OPP test.