background image

ProComSol, Ltd 

 

HM-BT-BAT User Manual 

 

MAN-1004 8/5/05

 

 

Page 5 

 

Troubleshooting 

 
Problem: 

Will not communicate 

 
Verify the following: 
1. Com port number in application is the HM-BT-BAT com port number. 
2. Loop power supply is on. 
3. Loop resistance between 250 ohms and 1Kohms. 
4. Loop current within HART limits. 
5. If multi drop configuration, all transmitters in loop have unique addresses. 
6. HM-BT-BAT HART connections across loop resistor or across transmitter terminals. 
7. Battery is fresh. 
8. Modem power switch is on and LED is illuminated. 
9. Perform the “Discovery” procedure again and verify a connection can be made. 
10. If using the Linksys USBBT100, verify the Linksys Bluetooth driver is installed and 
not the Windows Bluetooth driver.  There is a known issue with the Linksys install and 
Windows XP SP2.  Go to 

www.linksys.com

 and search for “USBBT100” for details. 

 
Problem: 

Communications unreliable 

 
Verify the following: 
1. You are in radio range of the master transmitter.  For Class 1 devices 100 ft, for Class 
2 devices, 25 ft. 
2. Vary the orientation of the master transmitter or the HM-BT-BAT to improve radio 
link strength. 
3. Battery is fresh. 
4. HART connections made before power turned on. 
5. Transmitter not in Burst mode.  Communications can occur in Burst mode, but more 
retries will be necessary for success. 
6. In some applications, a connection can be lost, which looks like a communication lock-
up.  Perform the Discovery process again to reestablish the link without the need to 
restart your application. 
 
Problem: 

Will not communicate with Cornerstone 

 
Perform the following: 
1. In directory “CSCONFIG/DB”, open file “CSLOCAL.INI”. 
2. Search for “[RDLS2]” without the quotes. 
3. Change “Debug=0” to “Debug=8”, again without the quotes. 
4. Save the file. 
 

Содержание HART HM-BT-BAT

Страница 1: ...etooth Range Note that Class 1 Bluetooth transmitters such as the USBBT100 have a range of 100 Class 2 transmitters such as the HP IPAQ PDA have a range of 25 The HM BT BAT works with both Class 1 and Class 2 transmitters but operation is limited to the range of the master transmitter Software Drivers No additional software drivers need to be installed Battery The HM BT BAT requires a standard 9V ...

Страница 2: ...V SPP and say it is not connected 6 Double click on ZV SPP The Bluetooth connection will be made and the assigned COM port will be reported This is typically COM8 Note this port number for your application software 7 Some Bluetooth drivers may prompt for a Passcode Enter 1234 without the quotes Discovery needs to be repeated only when adding or changing HM BT BAT modems or when multiple modems are...

Страница 3: ...or a Passcode Enter 1234 without the quotes Discovery needs to be repeated only when adding or changing HM BT BAT modems or when multiple modems are in the Bluetooth area Pocket PC Applications Bluetooth serial modems are generally assigned COM8 for application software use Start your PC application and set the com port setting to COM8 Use the application as normal The HM BT BAT looks like a norma...

Страница 4: ...o a HART device The program sends HART Command 0 to determine what transmitters are connected to the loop The Status box will indicate successful operation of the HM BT BAT in your system Consult the Troubleshooting section of this manual if test failure Pocket PC Test Software Program HM Test CE is included on the installation CD to test the operation of the HM BT BAT 1 Use the ActiveSync utility...

Страница 5: ...e Linksys install and Windows XP SP2 Go to www linksys com and search for USBBT100 for details Problem Communications unreliable Verify the following 1 You are in radio range of the master transmitter For Class 1 devices 100 ft for Class 2 devices 25 ft 2 Vary the orientation of the master transmitter or the HM BT BAT to improve radio link strength 3 Battery is fresh 4 HART connections made before...

Страница 6: ...cordance the instructions may cause harmful interference to radio communications However there is no guarantee that interference will not occur in a particular installation This device complies with part 15 of the FCC Rules Operation is subject to the following two conditions 1 This device may not cause harmful interference and 2 this device must accept any interference received including interfer...

Отзывы: