the “Cleaning and disinfection of the accessories” 
section.

Cleaning and maintenance

 

 

Cleaning of the compressor

Use a clean dry cloth and mild detergents to clean the 
outside of the unit.

ATTENTION: During cleaning, make sure the internal 
parts of the unit do not come in contact with liquids 
and the power cord is not connected.

Cleaning and disinfection of the accessories after 
each treatment:

The nebulizer can be cleaned by washing in liquid soap 
and water and then disinfected by boiling in water (max. 

o

10 minutes) or in an autoclave (max. 121 C for 10 
minutes). It is suggested that the mask, T-connector and 
mouthpiece be disinfected using cold disinfecting liquids. 
Dry the nebulizer using a clean dry cloth. After drying, 
connect it to the air outlet and switch the compressor ON 
to verify that it is functioning after cleaning.

Do not boil or autoclave the T-connector, 
mouthpiece, tubing or masks
It is important that the cleaning instructions are 
carefully followed to ensure optimal operation.

Replacement of the nebulizer.

The nebulizer should be replaced after a long period of 
inactivity, or if it is obstructed by dry medication, dust etc. 
With normal use, your nebulizer should last at least 6 
months.

Replacement of the filter

The filter is to be replaced after approximately 30 hours 
of use, or when it appears dirty. Open the filter holder to 
replace the filter with a new one. 
Do not use the device without the filter.

Troubleshooting
The device does not turn on:

1. Make sure the plug is firmly inserted into the electrical 
outlet.
2. Make sure the device has been operating within 
operating limits (30 minutes ON/ 30 minutes OFF)

• Use the accessories recommended by your doctor for 
therapy
16. Verify that the medication prescribed by your doctor 
can be used with common compressor nebulizer 
systems for aerosol therapy.

Instructions for use

Examine the unit before the first use in order to check for 
possible damage. 

1. When taking treatment with the device sit upright and 
relaxed.
2. After unpacking the device, check the compressor and 
accessories for visible damage or defects such like 
cracks in the plastic housing which could expose 
electrical components.
3. 

Before using the device,

 proceed with the cleaning 

operations as described in 

“Cleaning and disinfection 

of the accessories”

 section

4. Open the nebulizer by turning the top counter 
clockwise.
5. Make sure the baffle is positioned correctly inside the 
nebulizer chamber.
6. Put the prescribed quantity of medication into the 
medicine cup. Maximum fill volume is 6 ml.
7. Close the nebulizer by holding the bottom and turning 
the top clockwise making sure that the unit is properly 
closed.
8. Connect one end of the air tube to the nebulizer and 
the other end to the air outlet on the device.
9. Connect the mouthpiece to the nebulizer via the T-
connector.
10. If using the mask, remove the T-connector & 
mouthpiece and connect the mask directly to the tubing.
11. Plug the compressor into the electrical outlet making 
sure the electrical rating shown on the rating plate on the 
bottom of the unit corresponds to your main voltage.
12. To start the treatment, switch the device on by 
pressing the ON/OFF by setting the switch position to “I”.

ATTENTION: 

This device is rated for intermittent use 

with 30 minutes ON/ 30 minutes OFF.
13. When the treatment has been completed, switch the 
unit OFF by setting the switch position to “O” and 
disconnect the plug from the electrical outlet.
14. Wash the nebulizer and accessories as described in

The device does not nebulize, or the nebulization is 
weak:

1. Make sure the ends of the air tube are fitted tightly 
onto the main unit and onto the nebulizer.
2. Check whether the nebulizer is empty of 
medication or overfilled with medication (maximum fill is 
6ml)
3. Check whether the nebulizer nozzle is obstructed.
4. Check whether the baffle is inserted properly.

Note:
Never open the device. 

There are no serviceable parts 

within the unit. The compressor requires no lubrication or 
maintenance.