Proactive - TheraHeat
29
Proactive - TheraHeat
Electromagnetic Compatibility
Table 1 -
For all MEDICAL ELECTRICAL EQUIPMENT and SYSTEMS
Electromagnetic Compatibility
Table 2 -
For all MEDICAL ELECTRICAL EQUIPMENT and SYS
TEMS
Guidance and manufacturer’s declaration - electromagnetic immunity
This monitor is intended for use in the electromagnetic environment specified below.
The customer or the user of this monitor should assure that it is used in such an environment.
Immunity Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic Environment Guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material, the
relative humidity should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 6 kV contact
± 8 kV air
± 2 kV for power
supply lines
Main power quality should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
± 1 kV line(s) to
line(s)
± 2 kV line(s) to
earth
Main power quality should be that of a typical
commercial or hospital environment.
RF emissions
CISPR 11
Class B
This monitor is suitable for use in all
establishments, including domestic
establishments and those directly connected
to the public low-voltage power supply
network that supplies buildings used
for domestic purposes.
Содержание TheraHeat 715-440
Страница 1: ...Projet Instructions Please read this manual thoroughly before using this device for the first time...
Страница 37: ...Proactive TheraHeat 35 Notes...
Страница 38: ...Proactive TheraHeat 36 Notes...
Страница 39: ...Proactive TheraHeat 37 Notes...