![Prism Medical UK Mackworth Essentials 180 Скачать руководство пользователя страница 10](http://html1.mh-extra.com/html/prism-medical-uk/mackworth-essentials-180/mackworth-essentials-180_user-manual_1603508010.webp)
4.0 Final Inspection
Before first operation of the hoist:
• Confirm all cables are located and secure.
• Ensure the red emergency stop button on the control box is in the
out position (rotate clockwise to release if required).
• Press the up button on the handset and confirm the actuator
raises the boom.
• Press the down button on the handset and confirm the actuator
lowers the boom.
• Press the button on the handset to operate the leg opening and
confirm the actuator moves the legs outward.
• Press the button on the handset to operate the leg closing and
confirm the actuator moves the legs inward.
• Press the red emergency stop button on the control box an check
that actuators do not operate until the button is returned to the
out position.
• Check the emergency lowering function, both mechanical and
electrical, work properly.
• Check the rear wheel brakes work properly.
• Check the battery pack is fully charged.
Your hoist is now ready to use.