Page:
18
Rev:
04.2014
A
‐
205
‐
Owner’s
Manual
Emergency Stop & Lowering:
Caution: The manual emergency lowering system should be used only if the lowering procedures de-
scribed in the previous section of the manual do not work. Should you have any concerns or questions
contact your local authorised Prism Medical UK Representative.
Caution: DO NOT use the hoist after the
manual lowering
mechanism has been used. The hoist must be
reset by a qualified hoist technician after use.
Contact your local authorised Prism Medical UK Service
Provider.
Emergency Stop:
Push the Emergency Stop button to cut all power on the Mobile hoist. To resume power, release the emer-
gency stop button by turning in the direction as indicated by arrows on the Emergency Stop button.
Electrical emergency lowering/raising:
The hoist boom is lowered or raised by pressing the UP and DOWN arrow button on the control box.
Electrical emergency leg adjustment:
The hoist legs are opened or closed by pressing the leg opening, and leg closing arrow button on the con-
trol box.
Mechanical emergency lowering:
In case of power failure, it is possible to mechanically lower a patient placed in A-205 hoist.
In order to lower the boom in an emergency situation, pull up the red lever located directly on the hoist’s
actuator. The boom will lower as you press up and hold the lever. Release the red lever once you have
lowered the boom to a safe position.
Emergency Stop Button
Emergency Lowering Button
Emergency Raising Button
Emergency Leg Closing Button
Emergency Leg Opening Button
Содержание A-205