Princeton PHV-400 Скачать руководство пользователя страница 2

PHV-400.DOC 

Description: 

The PHV-400 Programmable PMT High Voltage Power Supply is a negative 
power supply, operating in the range of 0 to –1250 

V

olts and designed for use 

with photomultiplier tubes.  It is programmable by an analog input voltage 
over the entire voltage range.  The programming voltage is compatible 
with the 

Acton Series

 SC-1 SpectraCard 16-bit digital readout system.  

The PHV-400 PMT power supply is also manually adjustable by a front 
panel control and digital display. 
 

Installation and Set-up 

The PHV-400 PMT power supply operates over a wide range of input AC line 
voltages and is supplied with a line cord to match this input voltage.  No switching 
or rewiring is required to change operation to 100, 115 or 230 VAC nominal line 
voltage. 
 
Turn off the AC power switch on the left rear of the PHV-400.  Connect the high 
voltage cable supplied from the MHV connector on the rear of the PHV-400 to the 
high voltage MHV connector on the photomultiplier housing.  Turn the front panel 
control fully counter-clockwise to prevent accidental damage to the photomultiplier 
tube.  Plug in the supplied AC line cord to the rear of the PHV-400 and the other 
end of this line cord to the AC outlet. 
 

Operation: 

The PHV-400 PMT power supply can be operated remotely by an analog voltage 
from the Acton 

Series 

SC-1 SpectraCard or from the front panel control.  Remote 

or manual operation is determined by the position of the REMOTE/LOCAL 
switch on the rear of the PHV-400.   With the switch in the up or REMOTE 
position, the output voltage is determined solely by the input programming 
voltage where +5 

V

olts DC produces an output voltage –1250 

V

olts DC 

with a linear output to 0 volts.  Note that with no connector in the rear 
programming jack, the output voltage is 0 volts with the REMOTE/LOCAL switch 
in REMOTE.  The front panel control has no effect with the REMOTE/LOCAL 
switch in REMOTE.  The front panel digital meter reads the correct output voltage 
in both modes, however.  With the REMOTE/LOCAL switch in the down position 
or LOCAL position, the output voltage is determined by the front panel mounted 

Отзывы: