12
LS 785 High Throughput Lens Spectrograph User Manual
6.
If the CCD detector has a shutter or is controlling a shutter, open the
Trigger/Shutter
expander and set
the
Shutter Mode
to
Always Open
.
7.
Click on the
View
tab on the
Experiment
workspace and then click on the
Preview
button to begin
continuous live data acquisition.
8.
Rotate the focusing micrometer until the best focus (image quality or spectral resolution) is obtained.
CAUTION:
Do not apply excessive force to the knob.
Note:
The data acquisition rate is determined by the Exposure Time (set on the Common Acquisition
Settings expander). A shorter exposure time will update the viewer more frequently.
9.
Lock the micrometer in place by rotating the locking nut clockwise.
10.
Stop data acquisition.
11.
Turn off the light source.
12.
If a shutter was set to
Always Open
, return its setting to
Normal
.
Wavelength Adjustment
For wavelength adjustments, the LS 785 includes a “WAVELENGTH
ADJUSTMENT” micrometer located on the rear of the housing. Please
note that we refer to Wavelength Adjustment as moving (rotating) the
grating to a new center wavelength position on the sensor.
Procedure:
1.
With the LS 785, CCD detector, and neon light source powered
on, start LightField.
2.
After the LS 785 and CCD detector are detected, move their icons
into the
Experiment Devices
area.
3.
If the CCD detector has a shutter or is controlling a shutter, open
the Trigger/Shutter expander and set the
Shutter Mode
to
Always
Open
.
4.
Click on the
View
tab on the
Experiment
workspace and then
click on the
Preview
button to begin continuous live data
acquisition.
5.
Open the
Spectrometer
expander and enter the
Center Wavelength
.
6.
Under the
Micrometer Values
heading, note the Wavelength micrometer setting displayed.
7.
Rotate the micrometer clockwise or counter-clockwise until the desired micrometer setting is reached. If
desired, the micrometer can be locked into position by rotating the locking nut 1/8 turn clockwise.
8.
Stop data acquisition.
9.
Turn off the light source.
10.
If a shutter was set to
Always Open
, return its setting to
Normal
.
Figure 2-8. Wavelength Adjustment
Micrometer
Содержание LS 785