14
Monochromator Wavelength Movement Commands (continued):
>NM
Similar to NM except it returns control to user immediately rather than waiting for
completion of monochromator wavelength move. Can be used with ?NM or
MONO-?DONE below. This command must be terminated with MONO-STOP listed
below. NOTE: Use the NM command when communication with the
monochromator during the scan is not required.
?NM
Returns present wavelength in nm to 0.01nm resolution with units nm appended.
E.G.
?NM
300.00 nm
MONO-?DONE
Used with >NM command to determine if monochromator has reached the destination.
Returns 0 if move is not complete, 1 if move is complete.
MONO-STOP
Stops the monochromator wavelength move after use of the >NM command.
NM/MIN
Returns present scan rate in nm/min to 0.01 nm/min resolution with units nm/min
appended.
E.G.
?NM/MIN
100.00 nm/min
Grating Control Commands:
GRATING
Places specified grating in position to the wavelength of the wavelength on the present
grating. Up to nine (9) gratings are allowed on three (3) turrets. This command takes
a grating number from 1 - 9. IMPORTANT NOTE: This command assumes that the
correct turret is specified by the TURRET command. For example, using grating
numbers 1, 4 and 7 will place the first grating on the installed turret into that
position and call up the parameters for the grating number specified.
?GRATING
Returns the number of gratings presently being used numbered 1 - 9.
?GRATINGS
Returns the list of installed gratings with position groove density and blaze.
The present grating is specified with an arrow.
TURRET
Specifies the presently installed turret or the turret to be installed.
E.G. If installing the second turret, issue the command 2 TURRET to insure using
the correct parameters.
?TURRET
Returns the correctly installed turret numbered 1 - 3.
Diverter Mirror Control Commands:
EXIT-MIRROR
Designates the exit diverter mirror to receive the diverter control commands. This
command is for HRS spectrographs that can accept two diverter mirrors (HRS-500
and HRS-750). HRS-300 spectrographs will accept this command but it is not
required.
Содержание HRS-300
Страница 5: ...5 HRS 300 Features ...
Страница 6: ...6 ...
Страница 7: ...7 ...
Страница 20: ...20 Appendix 2 HRS Outline Drawing ...
Страница 21: ...21 Appendix 3 Outline Drawing Slit Assembly ...
Страница 22: ...22 Appendix 4 Gratings and Accessories ...
Страница 23: ...23 ...
Страница 24: ...24 ...
Страница 25: ...25 ...
Страница 27: ...27 Appendix 6 CE Declaration of Conformity ...