2. SAFETY AND COMPONENT PLACEMENT
This chapter lists safety precautions for use when operating the VersaSTAT 3; please
read them. There are suggestions for system component placement and information on
possible RF interference and transient sensitivity.
2.1. Safety Considerations
2.1.1. Line Voltage Settings and Fuses
The VersaSTAT 3
has a Corcom™ Power Entry Module on the left side of the rear panel.
The Corcom module (pictured below) contains the power connector and two line fuses.
The VersaSTAT 3 automatically detects the correct line voltage and frequency, so
there is no line voltage changes required in the Corcom.
The VersaSTAT 3 is supplied with the line voltage fuses in place, as well as extra fuses
supplied with each unit. For fuse replacement:
For 100
– 120 Vac line voltages, use a 2.0 A T (slow-blow) fuse (Littlefuse® type
213, 5 x 20 mm or equivalent).
For 220 - 240 Vac line voltages, use a 1.0 A T (slow-blow) fuse (Littlefuse® type
213, 5 x 20 mm or equivalent).
Caution: Do not use makeshift fuses or short-circuit the fuse holders
To replace a fuse: