5 Polarization and Conoscopy |
5.2 Centering the Objectives
Primotech | Document 430055-9044-008 V.1.0; 03 / 2014
33
5.2 Centering the Objectives
When viewing a sample on the rotating stage, it is important that the objective is
located exactly above the center of rotation of the stage.
Therefore, the objectives of Primotech D/A POL and Primotech D/POL Conoscopy
can be centered. The centering procedure should also always be performed after
changing an objective.
Procedure
1
Move the object guide to place a distinctive object in the sample at the center
of the crosshair by moving the sample on the table.
2
Slowly rotate the stage until the object is at the furthest point from the
crosshair.