Cholesterol  test  strips  only:  After  inserting  the  strip, 

“F”  (Fe-

male)

 on the left and 

“M” (Male) 

on the right will appear on the 

display.  Press  the  corresponding  button  (“-  “  for”F”  or  “+”  for 

“M”) to select the gender. After about two seconds, 

“A” (before 

meal)

 on the left or 

“P” (after meal)

 on the right will appear on 

the display. Press the corresponding button (“- “ for”A” or “+” for 

“P”) to select the letter. The selected letter will remain two sec-

onds on the display. After that, the code and the blinking drop 

will appear on the display, indicating that the device is ready to 

take the blood sample.

•  Lance the outer side of the fingertip with the lancing device and 

a sterile lancet to obtain a large hanging drop of blood. Wipe off 

the first drop with the cotton ball or gauze and use the second 

hanging drop of blood to apply to the test strip.

• 

The meter displays a drop symbol on the screen window, indi-

cating that the strip is inserted and ready for the blood sample. Apply a large 

hanging drop of blood directly from the finger to the white sample application 

area of the strip. For measuring glucose*, apply a drop of blood on the end of 

the test strip. Let the strip aspirate the blood sample. Do not touch the appli-

cation area with the finger. The drop of blood must be applied to the test strip 

immediately after lancing the fingertip. Blood which is applied later may lead to 

inaccurate result. For cholesterol and triglycerides testing it is critical to wipe 

away the first drop and immediately apply the second drop.

• 

Read and record your result. If the displayed result does not match your state 

of the health or seems to be unusually high or low, check the meter function 

using a new test strip and a control solution (not provided in the starter pack-

age). If this check confirms proper functioning of the instrument, read again 

the proceeding instructions of this manual. Perform another measurement us-

ing a new test strip. If the new result also seems to be not plausible, consult 

your doctor.

• 

When measurement is complete, use the ejector key on the back of the meter 

to remove the used strip.

•  Press power button until the instrument powers off.

• 

Properly dispose of the used lancets and test strip according to your local laws 

and guidelines.

• 

Clean the instrument if necessary (see the section “Care and Maintenance” 

of this manual).

QUALITY CONTROL RECOMMENDATIONS

To ensure that meter is functioning properly, you should consider using control 

solutions for a function check. For each test parameter, separate control solu-

tions are available to purchase. A function check is performed in the same way as 

a regular measurement except that control solutions are used instead of blood. 

Make a habit of carrying out a function check in the following situations: 

• 

If you leave the test strips container opened for a long time, or if the container 

integrity was compromised

• 

If you drop the meter

• 

If you repeated a test and the result is still not plausible, higher or lower than 

expected

• 

When you opened a new lot of the test strip container

• 

When you want to check the performance of the meter and the test strip.

MEMORY

The meter is capable of storing up to 500 test results and lets you review results 

in order from the most recent (1) to the oldest (500). When you insert a Data-chip 

into the CH meter, you can recall previous results. While in the standby, press 

and release ‘+’ and ‘-‘ button to enter MEMORY mode. The average result will 

be displayed. If you press ‘+’ or ‘-‘ button the memory number can be displayed. 

If you hold down ‘+’ or ‘-‘ the meter scrolls quickly and displays memory. A re-

sult corresponding to the memory position will be shown when you release the 

button. To delete one of the stored results press the power button and ‘-‘ button 

simultaneously for one second. To clear all the test results stored, continue to 

press power and ‘-‘ buttons for five seconds.

+

-

+

-

Содержание PL300

Страница 1: ... DIAGNOSTIC DEVICE PRIMA SELF TESTING KIT TRIGLYCERIDES CHOLESTEROL GLUCOSE ENGLISH PL300 PL311 Made in Italy PRIMA Lab SA Via Antonio Monti 7 6828 Balerna Switzerland info primahometest com primahometest com ONLY FOR THE PRIMA IN 0344 ...

Страница 2: ...rent types of stripes In case of any questions please call customer service INDEX What s the Prima 2in1 3in1 system for 3 Contents of the packaging 4 The digital meter 5 Display and symbols 6 Setting up the meter for the first time 7 Inserting the batteries 8 Meter settings SET 9 Requirements to perform a measurement 11 Introduction to test strips and data chips 11 Programming the device 12 How to...

Страница 3: ...abetes hypertension hypercholesterolemia and family factors also linked to an unhealthy lifestyle such as smoking alcohol abuse and a diet rich in saturated fats in combination with a lack of physical activity In ad dition stress oral contraceptives and pregnancy can cause elevated cholesterol levels A high concentration of cholesterol is a major risk factor for cardiovascular events The Canadian ...

Страница 4: ...dicator 4 Test results and other messages 5 Units of measure 6 Parameter being tested 7 Temperature indicator 8 Time 9 Day Month Year 10 Beeper Meter s modalities The meter has three modalities STANDBY It s the default mode that appears every time you turn on the instrument by pressing the ON switch The strip symbol will flash and the display will show the number of the data chip the date and time...

Страница 5: ...2032 in the slot as shown paying at tention to the symbols and ends of the batteries Use only 3V lithium batteries CR 2032 Remember to always change both batteries at the same time as batteries with different capacity may interfere with the functionality of the instrument 3 Close the battery compartment by sliding the cover until you hear a click 4 Turn the instrument to test the operation of the ...

Страница 6: ...wer right corner To set the preferred format press and release the power button to access the SET until you see the format blink Press the or to select date formats To select the month day and year press and release the power button until the signal does not blink Press or buttons to set the correct month Press power to confirm and go to the next setting Similarly also set the date and year by pre...

Страница 7: ...or tests which require one or two drops of blood An ad CH REQUIREMENTS TO PERFORM A MEASUREMENT Your PRIMA Self Testing device 2in1 or 3in1 Test strips for measuring the desired parameter on chip data Lancing Device Sterile Lancet User s manual Cotton swab or gauze not included Diary where you record the results not included INTRODUCTION TO TEST STRIPS AND DATA CHIPS The basic package of PRIMA Sel...

Страница 8: ...g de vice by turning it counterclockwise while holding the base 3 While holding the lancet carrier insert a new sterile lancet into the bottom of lancet carrier Twist the cap of the lancet off 5 Hold the tip firmly in one hand then pull out the sliding barrel with the other hand This will cock the lancing device 2 With two fingers pull out the lan cet carrier and hold it in place 4 Put the tip of ...

Страница 9: ...tainer Close the container immediately after removing strip to protect the desiccants and re maining strips otherwise the test strips may become unusable before expiry date Liquids must not enter the test strip container Hold the test strip with thumb and index finger so that the blood sample appli cation area is facing upwards Insert the test strip into the designated strip slot From the printing...

Страница 10: ...the proceeding instructions of this manual Perform another measurement us ing a new test strip If the new result also seems to be not plausible consult your doctor When measurement is complete use the ejector key on the back of the meter to remove the used strip Press power button until the instrument powers off Properly dispose of the used lancets and test strip according to your local laws and g...

Страница 11: ...tton clean tissue or cloth Mild soap as well as 70 alcohol or isopropanol al cohol are suitable to moisten the cotton cloth to clean plastic surfaces Lift the front cover and gently clean glass surfaces of the optic with a dry gauze or cotton ball Clean plastic cover support of the meter with mois tened and then dry gauze or cloth Do not use any disinfectant sprays or cloth cotton balls which are ...

Страница 12: ...dling accident and or negligent care IMPORTANT NOTES Always Operate the instrument within the acceptable temperature range Place the instrument on a level surface or hold it steady in your hand Make sure that all display elements are functioning during the self test Read the test strip inserts Keep the meter and the test strip container clean and tidy To get less variable results fast at least 10 ...

Страница 13: ...cets 5 cholesterol strips 5 triglycerides strips 10 glucose strips PL311 Sterile lancets for Lancing device 50 Sterile lancets 100 Sterile lancets PL313 PL314 PRODUCT LIMITATIONS Please read the instructions of use for the test strips and control solutions for detailed information on product data and the respective limits of the test PRIMA Glucose test strips with the blue data chip and PRIMA Gluc...

Страница 14: ...25 26 LOGBOOK Write down the results you get with the PRIMA device ...

Отзывы: