![Preventice Technologies BodyGuardian HEART Скачать руководство пользователя страница 37](http://html1.mh-extra.com/html/preventice-technologies/bodyguardian-heart/bodyguardian-heart_instructions-for-use-manual_1599547037.webp)
31
Appendix B: Electromagnetic environment
Appendix B:
Electromagnetic environment
The BodyGuardian Heart monitor is suitable for use in the specified
electromagnetic environment. The purchaser or user of the monitor should
assure that it is used in an electromagnetic environment as described below:
Emissions
Emissions test
Compliance Electromagnetic Environment
RF emissions
CISPR 11
Group 1
The monitor uses RF energy only for its internal
function. Therefore, the RF emission is very low
and not likely to cause any interference in nearby
electronic equipment.
RF emissions
CISPR 11
Class B
The monitor is suitable for use in domestic
establishments and in establishments directly
connected to the low voltage power supply
network which supplies buildings used for
domestic purposes.
Harmonic
emissions
IEC 61000-3-2
Class A
Voltage
fluctuations/
flicker
emissions
IEC 61000-3-3
Complies