When using a Precision Medical Carry Bag, slide the cylinder into
the bag with the gauge facing the mesh. Attach the cannula to the
Conserver’s Cannula Connection.
Slowly open the cylinder valve. Turn Dial to the prescribed setting.
Zip the bag closed.
EasyPulse5 Oxygen Conserving Regulator (Conserver)
Dial
T-Handle
Contents
Gauge
(if equipped)
Cylinder
Index
Pins
Cannula
Connection
Post Valve
Index Pin
Alignment
Holes
4
3
5
6
1
2
P
A
R
T
S
S
T
E
P
S
T
E
P
S
T
E
P
S
T
E
P
S
T
E
P
S
T
E
P
I
d
When finished using:
Turn the cylinder valve to the OFF position. Turn Conserver Dial to “2 LPM Continuous”
position. Wait for Oxygen to stop flowing from Conserver. Remove Conserver from Post
Valve. Store Conserver and Carry Bag in a safe place.
Place cannula in nostrils
and breathe
through
nose to begin Oxygen
flow. If using a bag
adjust the straps for
maximum comfort.
Typical bag carrying configurations
Cannula prongs in
nostrils
Yoke Seal
View of properly
attached conserver
Inspect Conserver and Post Valve for damage or contaminants.
Do NOT use if any are found. “Crack” cylinder valve by opening the
valve slightly and closing.
Ensure Conserver is in the OFF position and Yoke Seal is in place.
Align
Index Pins with Index Pin Alignment Holes in Post Valve.
Hand tighten T-Handle. Do NOT use tools.
Carry
Bag
Cannula
Oxygen