PROBLEM
PROBABLE CAUSE
REMEdy
Precision Medical, Inc.
300 Held Drive • Northampton, PA 18067
800-272-7285 • 610-262-6090 • fax: 610-262-6080
504345rev3 25M 9/07 Printed in USA
A. No flow
1. Cylinder valve closed
1. Turn on cylinder
2. Conserver in “OFF”
2. Set to prescribed
position
setting
3. Cylinder empty
3. Replace cylinder
4. Dial not at setting
4. Set dial so indicator
points to a setting
5. Conserver not sensing
5. Check position of
breath
cannula in nose
Do not breathe
through mouth
B. Leaking at Yoke
1. Missing or defective
1. Replace Yoke Seal
Yoke Seal
If problem cannot be corrected, consult your dealer/Homecare equipment provider.