3

SPECIfICATIONS

dimensions:

 (Are approximate and may vary by Model)

 

Weight:

 

empty 

2.9 lbs (1.32 kg)

 

Full 

3.6 lbs (1.63 kg)

 

Full (including all accessories) 

4.6 lbs (2.09 kg)

 

length: 

4.81” (12.2 cm)

 

Width: 

3.63” (9.2 cm)

 

Height: 

8.19” (20.8 cm)

Operating Conditions:

 

Temperature: 

40°F to 110°F (4.4°c to 43.3°c)

 

Altitude: 

Sea level to 10,000 ft (3048 m)

Storage Conditions:
 

Temperature: 

-10°F to 140°F (-23°c to 60°c)

 

Humidity: 

95% Non condensing

Pulse Setting:

 

0, 1, 2, 3, 4, lPM equivalents

Maximum Capacity:

 

0.32 liters

Maximum Working Pressure:

  52.94 PSI (3.65 bar)

Cannula Requirement:

 

Maximum 7 foot long standard or high 
flow adult single lumen oxygen nasal 
cannula

Pulse Volume Accuracy:

 

Within ±15% of the nominal bolus value 
(at each breath rate)

Trigger Method:

 

Inspiratory effort (negative pressure 
from patient inhalation)

breathing frequency:

 

1 to 30 breaths Per Minute

Specifications subject to change without prior notice.

INTENdEd USE

To be used as a portable, supplemental, refillable oxygen device which 
delivers USP (United States Pharmacopeia) oxygen at a number of 
various pulsed settings.  The device is to be used by patients who may 
have difficulty extracting oxygen from the atmosphere. It is for patients 
who would normally receive oxygen via nasal cannula. It is intended to 
be used as an ambulatory source of oxygen.

This product is not intended as a life-sustaining or life-supporting 
device.

dANgER

Содержание EasyMate PM2200

Страница 1: ...ON SAVE THESE INSTRUCTIONS Federal USA law restricts this device to sale by or on the order of a physician Precision Medical Inc 300 Held Drive Tel 001 610 262 6090 Northampton PA 18067 USA Fax 001 61...

Страница 2: ...y bag 12 cleaning 13 Replacement parts 13 Maintenance 13 Returns 13 TROUBLESHOOTING 14 DISPOSAL 15 NOTES 16 WARRANTY 17 DECLARATION OF CONFORMITY 18 Receiving inspection Remove product from package an...

Страница 3: ...RMENTS WILL PRODUCE OXYGEN ENRICHED ATMOSPHERE Symbol for SINGLE USE Applies to Cannula only 2 Symbol for Warning low temperature freezing conditions Symbol for Warning High pressure oxidizing gas vig...

Страница 4: ...ow adult single lumen oxygen nasal cannula Pulse Volume Accuracy Within 15 of the nominal bolus value at each breath rate Trigger Method Inspiratory effort negative pressure from patient inhalation Br...

Страница 5: ...flames Use and store Reservoir and Portable at least five feet away from equipment such as furnaces water heaters and stoves that may contain open flames Keep equipment in a well ventilated area at a...

Страница 6: ...it is recommended that the Portable Liquid Oxygen System be qualified on patients in the situations it will be used rest exercise sleep This device is designed to operate with a single lumen adult na...

Страница 7: ...g the various pulse settings The device should be filled just before use Fill Connector Appearance may vary per model Vent to Fill Lever Pulse Selector w Label PORTABLE system description Cannula Conn...

Страница 8: ...WHEN LOCKING INTO POSITION PM2202 Caire Side Fill Rotate the Portable System counterclockwise until the pin of the device engages with the slot of the reservoir connector approximate rotation 45 Care...

Страница 9: ...and when attempting to disconnect it Figure 2 PM2201 PM2202 Disconnect the Portable System from the Reservoir by rotating it in counterclockwise direction until the device separates The device may now...

Страница 10: ...rk DO NOT USE FORCE Simply allow a few minutes for the frozen parts to warm then disengage the Portable when the ice has melted 11 Check the approximate oxygen contents in the Portable System using th...

Страница 11: ...the contents scale and allow it to snap back into the scale doing so will damage the scale and void the warranty The table below shows approximate use times for the Portable System after it has been...

Страница 12: ...vice to occasionally skip a breath or to pulse twice in one breath 5 Breathe through the nose and feel a pulse of oxygen as you inhale 6 Insert Portable System into the carry bag Carry bag may become...

Страница 13: ...Placing the device on its side or upside down will shorten the usage time device Figure 6 Using the Carry Bag Figure 7 illustrates how the device should be placed in its bag Ensure the device is orien...

Страница 14: ...ent Replacement parts Description Part Cannula 504833 Carry Bag Liner 504488 Carry Bag with Liner 504392 Contents Scale 504393 MAINTENANCE ATTENTION Equipment Provider This device contains several fie...

Страница 15: ...n 1 Refill device 2 Set to prescribed setting 3 Rotate Pulse selector to your prescribed setting 4 a Check position of cannula in nose b Do not breathe through mouth 5 Reconnect cannula 6 Remove kink...

Страница 16: ...Allow time for device to warm D Device frosted no pulse 1 Device overfilled 1 Wait approximately 30 minutes until unit returns to normal operating conditions E Device is making a hissing noise venting...

Страница 17: ...16 NOTES device serial In Service Date Date...

Страница 18: ...ions have been made to the goods shall in its discretion and at its own expense repair or replace the defective component s ORAL STATEMENTS DO NOT CONSTITUTE WARRANTIES AND THERE ARE NO WARRANTIES OTH...

Страница 19: ...spiratory Therapy Equipment Part 2 Tubing And Connectors 2002 EN 1418 Welding personnel approval testing of welding operators for fusion welding and resistance weld setters for fully mechanized and au...

Страница 20: ...n USA ISO 13485 Certified Precision Medical Inc 300 Held Drive Northampton PA 18067 USA Visit us on the Web www precisionmedical com Toll Free Tel 800 272 7285 Toll Free Fax 800 353 1240 Tel 001 610 2...

Отзывы: