Precision Medical Easy Pulse POC PM4100 Series Скачать руководство пользователя страница 17

16

Portable Oxygen Concentrator

ALARM / INDICATORS / TROUBLESHOOTING

PM4150

Visual Indicator/

Problem

Audible Alarm

Probable Cause

Remedy 

Red alarm indicator 

is illuminated, unit 

is not running and 

setting numbers 1 & 

3 are flashing

Repetitive beep 

Internal battery is 

outside its operating 

temperature range

1. Turn off the POC and disconnect any 

power supplies

2. Check that the inlet filter is clean

3. Ensure inlet and exhaust vents are not 

blocked in any way

4. Verify you are operating the POC within 

the specified operating temperature 

range.

5. It may be necessary to allow the POC 

to warm up or cool before turning back 

“ON”

Red alarm indicator 

is illuminated, unit 

is not running and 

setting numbers 1 & 

4 are flashing

Repetitive beep 

Motor is above its 

maximum operating 

temperature

1. Turn off the POC

2. Check that the inlet filter is clean

3. Ensure inlet and exhaust vents are not 

blocked in any way

4. Verify you are operating the POC within 

the specified operating temperature 

range

5. It may be necessary to allow the POC to 

cool before turning back “ON”

Red alarm indicator 

is illuminated, unit 

is not running and 

setting numbers 1 & 

5 are flashing

Repetitive beep

Pressure outside of 

normal operation

1. Select the “OFF” position

2. Select the prescribed setting

3. If alarm persists, change to alternate 

oxygen source and contact your 

Equipment Provider

Red alarm indicator 

is illuminated, unit 

is not running and 

setting numbers 2 & 

3 are flashing

Repetitive beep 

Fan malfunction

1. Select the “OFF” position

2. Select the prescribed setting

3. If alarm persists, change to alternate 

oxygen source and contact your 

Equipment Provider

Red alarm indicator 

is illuminated, unit 

is not running and 

setting numbers 2 & 

4 are flashing

Repetitive beep 

Failed electrical 

connection

1. Select the “OFF” position

2. Select the prescribed setting

3. If alarm persists, change to alternate 

oxygen source and contact your 

Equipment Provider

Amber alarm 

indicator is flashing 

and unit is running

NA

Service Needed

1. Continue to use POC

2. Contact Equipment Provider

Red alarm indicator 

is flashing and unit 

is running

NA

Service Needed

1. Switch to another source of oxygen 

2. Contact Equipment Provider

Red Alarm and 3 & 

4 Setting Numbers 

Illuminated and 

device Shuts Down

Single Beep

Cannula Occluded/

Kinked

1. Ensure the Cannula and/or Outlet are 

Not Blocked

2. Select the “OFF” position and wait at 

least 15 seconds.

3. Select the prescribed Setting

4. If Alarm persists, change to alternate 

Oxygen source and Contact your 

Equipment Provider.

Содержание Easy Pulse POC PM4100 Series

Страница 1: ...ies SAVE THESE INSTRUCTIONS Federal USA law restricts this device to sale by or on the order of a physician 300 Held Drive Tel 001 610 262 6090 Northampton PA 18067 USA Fax 001 610 262 6080 ISO 13485...

Страница 2: ...SAL INSTRUCTIONS 17 LIMITED WARRANTY 18 DECLARATION OF CONFORMITY 19 RECEIVING INSPECTION Remove the Precision Medical Inc EasyPulse Portable Oxygen Concentrator POC from the packaging and inspect for...

Страница 3: ...t the safety alert symbol indicates a potentially hazardous situation which if not avoided may result in property damage 62NA PM4130 AND PM4150 WITH RESPECT TO ELECTRICAL SHOCK FIRE AND MECHANICAL HAZ...

Страница 4: ...DO NOT reach for the POC that has fallen into water Unplug immediately DO NOT attempt to repair POC Any attempt could result in electrical shock POC is designed to be used in dry conditions the POC is...

Страница 5: ...NOT place the POC in a small closed space such as a closet DO NOT use extension cords or electrical adapters The POC is designed to operate with a single lumen adult cannula with a maximum length of...

Страница 6: ...here the oxygen concentrator is located If unable to leave the room you must wait 10 minutes after you have turned the oxygen concentrator off before smoking Open flames during oxygen therapy are dang...

Страница 7: ...equipment provider are accountable for ensuring the compatibility of the oxygen concentrator and all of the parts or accessories used to connect to the patient before use The Oxygen Concentrator its...

Страница 8: ...39 dBA PM4150 44dBA Cannula Maximum 7 ft 2 1 m long standard adult single lumen nasal cannula Equipment Classification Class II with respect to protection from electrical shock Type BF degree of prote...

Страница 9: ...Additional monitoring or attention may be required for patients using this POC who are unable to hear or see alarms or communicate discomfort The Setting Knob does not rotate 360 Rotating the knob le...

Страница 10: ...er Adapter DC Power Adapter or Internal Battery Refer to the POWER SUPPLY section of the user manual for details on these 3 options 6 Turn ON the POC by selecting the setting for your prescribed flow...

Страница 11: ...e carry bag is not intended to make the POC waterproof or to protect the device from major falls or accidents abuse When inserting the POC into the carry bag make sure of the following Before placing...

Страница 12: ...he POC cool and operating properly Be sure the exhaust vents are facing away from body while being carried Iftheoutsideofthebagfeelshotorcausesanydiscomfort set the POC down and check to be sure all v...

Страница 13: ...rnal power sources AC Power Adapter The AC power adapter connects the POC to a 100 240 VAC 50 60 Hz wall outlet The power adapter converts AC voltage to DC voltage in order to power the POC The AC ada...

Страница 14: ...and simultaneously charge the internal battery DC Plug Auto Accessory Plug Auto Accessory Plug DC Plug connected to Power Input Auto 12 V Accessory Jack Connection To connect the DC Power Adapter to t...

Страница 15: ...ternal battery will begin charging as soon as power is connected You will also be able to run the POC while the internal battery charges Lithium Ion Battery Guidelines Proper Use In order to prolong l...

Страница 16: ...o breath detected for 2 minutes 1 Verify the cannula is not kinked 2 Verify the cannula is properly connected to oxygen outlet 3 Verify the cannula is properly positioned in the nose 4 If alarm persis...

Страница 17: ...itive beep Pressure outside of normal operation 1 Select the OFF position 2 Select the prescribed setting 3 If alarm persists change to alternate oxygen source and contact your Equipment Provider Red...

Страница 18: ...e POC until it is thoroughly dry 7 When not in use store the POC in a clean area free from grease oil and other sources of contamination Inlet Filter 1 Remove the inlet filter by pulling on the protru...

Страница 19: ...n expense ORAL STATEMENTS DO NOT CONSTITUTE WARRANTIES The representatives of Precision Medical Inc or any retailers are not authorized to make oral warranties about the merchandise described in this...

Страница 20: ...relevant provisions of the aforementioned legislation and the result entitles the organization to use the CE 0473 marking on those products listed above Applied Standards EN 1041 EN ISO 14971 EN 6060...

Отзывы: