Canister Removal/Assembly (1200cc Reusable)

1. Grasp filter and pull apart

3. Remove lid, empty & clean

5. Align indicators on lid and canister

7. Align with port & push together

2. Tilt off base to remove

4. Place lid on canister

6. Tilt and place canister on device base

8. Correctly installed canister is level

Содержание Easy Go Vac PM66 Series

Страница 1: ...Shown 300 Held Drive Tel 001 610 262 6090 Northampton PA 18067 USA Fax 001 610 262 6080 ISO 13485 Certified www precisionmedical com SAVE THESE INSTRUCTIONS Read this manual before operating the devic...

Страница 2: ...g the Instructions of Use written in the appropriate language This manual is provided for your training safety and to prevent damage to the Aspirator Ifyoudonotunderstandthismanual DONOTUSEtheAspirato...

Страница 3: ...ale by or on the order of a physician This device may contain electrical components that are hazardous to the environment DO NOT dispose device into standard trash Contact your local waste Management...

Страница 4: ...ing electromagnetic interference with other devices when used normally in accordance with their intended use Should the aspirator cause any interference try repositioning it or moving it to another lo...

Страница 5: ...on AC power PM66Li 50 to 550 mmHg 1 96 to 21 6 inHg PM66 200 10 to 210 mmHg 0 394 to 8 26 inHg Electrical Car Charger DC 12 Volts 1 8 Amps Household Current AC 100 240 Volts 50 60 Hz 0 60 0 25 Amps 6...

Страница 6: ...0 4 4 2kV on AC Mains 2kV on AC Mains Mains power quality should be that of a typical commercial or hospital environment Surge IEC 61000 4 5 1kV Differential 1kV Common 1kV Differential 1kV Common Mai...

Страница 7: ...t electromagnetic interference by maintaining a minimum distance between portable and mobile RF Communications Equipment and the device as recommended below according to the maximum output power of th...

Страница 8: ...ter Built in overfill protection device 800ml Suction Canister Built in overfill protection device Vacuum Gauge Power Source Indicator Charge Status Indicator Battery Level Indicator Exhaust Vents On...

Страница 9: ...requires charging CHARGING The power switch is OFF AC is plugged in and light is AMBER the Aspirator is in the charging process FULL The power switch is OFF AC is plugged in and light is GREEN the Asp...

Страница 10: ...end to set vacuum level Push button Operating Instructions Plug into wall outlet DC outlet Power OFF Full charge 2 to 3 hrs Push together Ensure all connections are tight 6 Perform Procedure Connect p...

Страница 11: ...rted until the battery has been recharged or until the Aspirator is connected to either of the external power source options AC or DC WARNING Operating Aspirator after Amber Low Battery Light has been...

Страница 12: ...d set aside NOTE Filter MUST NOT GET WET The filter material cannot be cleaned 2 Remove canister lid wash in warm water dishwashing solution Rinse with clean warm water 3 Soak canister and lid in 1 pa...

Страница 13: ...submerge in water as this will result in damage to the vacuum pump NOTE Do not use any cleaners or disinfectants that contain ammonia benzene and or acetone to clean the device Suction Device multi p...

Страница 14: ...d empty clean as instructed in Cleaning Maintenance section 5 Place canister on base 7 Push together until they engage click 2 Pull canister away from device 4 Reinstall lid securely 6 Align canister...

Страница 15: ...p filter and pull apart 3 Remove lid empty clean 5 Align indicators on lid and canister 7 Align with port push together 2 Tilt off base to remove 4 Place lid on canister 6 Tilt and place canister on d...

Страница 16: ...and says OUT is facing toward inlet port on device Filter cannot be cleaned If it becomes contaminated or clogged it must be replaced Slightly twist and pull up Elbow Elbow Inline Filter OUT toward d...

Страница 17: ...e as described by your Healthcare Professional or Healthcare Facility protocol WARNING The Aspirator may contain an internal battery Batteries contain materials which can contaminate the environment w...

Страница 18: ...On Off Switch to ON position 2 Turn on car to activate it s DC outlet 3 Check connections at both ends of cord 4 Connect to a known working auto power port 5 Contact your Provider Battery will not cha...

Страница 19: ...DC Power Cord Carry Bag Only Description Part Inlet Filter Hydrophobic 506851 Inlet Filter Bacterial 507754 800cc Canister Case 10 ea 502519 10 800cc Inlet Adapter Kit 507521 1200cc Canister 507522 P...

Страница 20: ...Inc oranyretailersare not authorized to make oral warranties about the merchandise described in this contract and any such statements shall not be relied upon and are not part of the contract for sale...

Отзывы: