PD6800 Loop-Powered Process Meter Instruction Manual
36
Maximum & Minimum Readings (
MAXIMUM
&
MINIMUM
)
The maximum and minimum (peak & valley) readings reached by the process are
stored in the meter since the last reset or power-up. The meter shows
MAXIMUM
or
MINIMUM
to differentiate between run mode and max/min display. Press
Enter
to
remain in Max/Min display mode. If
Enter
is not pressed, the Max/Min display
readings will time out after ten seconds. The meter will return to display the actual
reading.
4065.3
MAXIMUM
3024.7
GALLONS
Run Mode
2457.7
MINIMUM
Press
Up
to Display and to
Toggle Between Max & Min
10 Sec Time Out
Press
Enter
to hold Max/Min
Press
Right
to Reset Max/Min
Press
Menu
to Exit Max/Min
Reset Meter to Factory Defaults
When the parameters have been changed in a way that is difficult to determine
what’s happening, it might be better to start the setup process from the factory
defaults.
Instructions to load factory defaults:
Enter the
Advanced
features menu.
Press and hold
Reset
button when
INFO
is shown.
Press
Enter
when
RESET
DFALTS?
prompt is shown
Note: If
Enter
is not pressed within three seconds, the prompt will stop flashing return
to showing
INFO
.
INFO
RESET
DFALTS?
LOADING
Press and
Hold for 5 sec
Flashing
Display
Defaults
Restored