Javelin D
Ɣ
Model PD644 Voltmeter
Instruction Manual
32
Scaling the 4-20 mA Analog Output (
Aout
)
The 4-20 mA analog output can be scaled to provide a 4-20 mA signal
for any display range selected.
No equipment is needed to scale the analog output, simply program the
display values to the corresponding mA output signal.
The
Analog Output
menu is used to program:
1. 4-20 mA output based on display values
2. Sensor break value in mA
out1
0
4.00
Press Enter to Accept Setting
dis2
Scale
Output 2
dis1
Set Display 1
Value
scal
0
0
00
Set Output 1
Value
Press Up to Set Digit Value
Press Right to Select Next Digit
Press Menu to Exit at any Time
Aout
Analog Output when Display is Out of Range
The analog output reflects the display out of range conditions as follows:
Input Condition
Display
Analog Output
Underrange Flashing
-1999
3.00
mA
Overrange Flashing
9999
21.00
mA