PD662 Loop-Powered Meter
Instruction Manual
6
SPECIFICATIONS
Except where noted all specifications apply to operation at +25°C.
General
DISPLAY
0.6" (15.24 mm) LCD, 3½+ digits; -1999 to 2999
DISPLAY
UPDATE RATE
2 Updates/Second
OVERRANGE
Display flashes
2999
UNDERRANGE
Display flashes
-1999
PROGRAMMING
METHOD
Four internal pushbuttons
NOISE FILTER
Programmable
HI
,
LO
, or
OFF
RECALIBRATION
Recalibration is recommended at least every 12 months.
MAX/MIN
DISPLAY
Max/Min readings reached by the process are stored until
reset by the user or until power to the meter is turned off.
NON-VOLATILE
MEMORY
All programmed settings are stored in non-volatile memory
for a minimum of ten years if power is lost.
NORMAL MODE
REJECTION
64 dB at 50/60 Hz
ENVIRONMENTAL
Operating temperature for CSA certification: -40 to 75°C
Functional temperature range: -40 to 85°C
Storage temperature range: -40 to 85°C
Relative humidity: 0 to 90% non-condensing
CONNECTIONS
Removable screw terminals accept 12 to 22 AWG wire
ENCLOSURE
Impact resistant polyester plastic, body gray, cover clear
with blue faceplate; NEMA 4X, IP66; ½" conduit hole
provided at base.
MOUNTING
½” NPT pipe (0.865 in, 12.7 mm) conduit hole on bottom
of enclosure. 0.166 in (4.2 mm) wall mounting holes
located behind front cover screws.
TIGHTENING
TORQUE
Screw terminal connectors: 4.5 lb-in (0.5 Nm)
OVERALL
DIMENSIONS
4.33" x 3.15" x 2.76" (110 mm x 80 mm x 70 mm)
4.33" x 3.15" x 3.35" (110 mm x 80 mm x 85 mm)
for models with heater option.
WEIGHT
10.4 oz (295 g)
WARRANTY
3 years parts and labor