background image

PD6820 Explosion-Proof Loop-Powered Rate/Totalizer 

Instruction Manual 

 

12

 

Main Menu Display Functions &  

Messages 

The meter displays various functions and messages 

during setup, programming, and operation. The table 

in the right column shows the main menu functions 

and messages in the order they appear in the menu. 

 

Main Menu 

The main menu consists of the most commonly used 

functions: 

Setup, Password, 

and 

Service

Press 

MENU

 button to enter Programming Mode then 

press the 

Up

 

Arrow

 button to scroll through the main 

menu. 

 

  Press 

MENU

, at any time, to exit and return to 

Run Mode. Changes made to settings prior to 

pressing 

Enter 

are not saved. 

  Changes to the settings are saved to memory 

only after pressing 

Enter

  The display moves to the next menu every time  

a setting is accepted by pressing 

Enter

 

Display 

Parameter 

Action/Setting 

SETUP

 

Setup 

Enter 

Setup

 menu 

DeC..pt

 

Decimal point 

Enter 

Decimal Point 

menu 

Rate

 

Rate decimal 

Set rate display decimal 

point 

total

 

Total decimal 

Set total display decimal 

point 

PRoG

 

Program 

Enter the 

Program

 menu 

sCalE

 

Scale 

Enter the 

Scale

 menu 

Cal

 

Calibrate 

Enter the 

Calibrate

 menu 

Inpt1

 

Input 1 

Calibrate input 1 signal or 

program input 1 value 

DspL1

 

Display 1 

Program display 1 value 

Inpt2

 

Input 2 

Calibrate input 2 signal or 

program input 2 value 

DsPl2

 

Display 2 

Program display 2 value 

Span 

Error

 

Span 

Error 

Error, calibration not 

successful, check signal 

TbASE 

Time Base 

Enter the 

Time Base 

menu 

sec 

Second 

Units per 

second

 

m

i

Minute 

Units per 

minute

 

hour 

Hour 

Units per 

hour

 

day 

Day 

Units per 

day

 

Tot

C

Conversion 

Factor 

Enter the 

Conversion Factor 

menu 

T rST

 

Total Reset 

Enter the 

Total Reset 

menu 

Auto

 

Automatic 

Automatic

 Total Reset 

T DELAY 

Time Delay 

Automatic Reset 

Time 

Delay

 

man

 

Manual 

Manual

 Total Reset 

Enabl 

Enable 

Enable

 Manual reset

 

dsabl 

Disable 

Disable

 Manual reset

 

tag

 

Tag/Units 

Enter the Tag/Units Menu

 

ON

 

Tag On 

Enable Tag/Units

 

OFF

 

Tag Off 

Disable Tag/Units

 

togle

 

Tag Toggle 

Toggle Tag and Total

 

PASSWRD

 

Password 

Enter the 

Password

 menu

 

UNLOCKD

 

Unlocked 

Program password to lock 

meter

 

LOCKED

 

Locked 

Enter password to unlock 

meter

 

99999 
-99999

 

Flashing 

display 

Overrange condition 

Underrange condition

 

SERVICE

 

Service 

Select before 

removing/installing cover for 

service or to clean the glass 

window

 

 

 

 

Содержание PD6820

Страница 1: ...ammable Exponent Feature Open Collector Pulse or Alarm Output Display Mountable at 0 90 180 270 SafeTouchTM Through Glass Button Programming HART Protocol Transparent Loop or External DC Powered Backlight Standard Operating Temperature Range 40 to 75 C 40 to 167 F FM Approved as Explosion Proof Dust Ignition Proof Flame Proof CSA Certified as Explosion Proof Dust Ignition Proof Flame Proof ATEX an...

Страница 2: ...ective unit See Warranty Information and Terms Conditions on www predig com for complete details 2020 Precision Digital Corporation All rights reserved Introduction The ProtEX RTA PD6820 is a rugged explosion proof loop powered rate totalizer fully featured for demanding applications in hazardous areas or in the harshest environmental conditions The meter derives all of its power from the 4 20 mA ...

Страница 3: ...s Messages 12 Main Menu 12 Setting Up the Meter SETUP 13 Setting Numeric Values 13 Setting the Decimal Point dec PT 13 Programming the Meter Prog 14 Setting the Time Base t base 15 Setting the Total Conversion Factor totCF 15 Manual or Automatic Total Reset Function t rST 16 Setting the Tag Display tag 16 Setting Up the Password PASSWRD 17 Locking the Meter 17 Making Changes to a Password Protecte...

Страница 4: ...23 Quick User Interface Reference 24 EU Declaration of Conformity 25 Table of Figures Figure 1 Enclosure Dimensions Front View 7 Figure 2 Enclosure Dimensions Side Cross Section View 7 Figure 3 Connector Board 8 Figure 4 Connections without Backlight 9 Figure 5 Connections with Loop Powered Backlight 9 Figure 6 Connections with Externally Powered Backlight 9 Figure 7 Reset Connections 9 Figure 8 C...

Страница 5: ...y be mounted directly to conduit Two slotted flanges for wall mounting or NPS 1 to 2 or DN 40 to 65 mm pipe mounting See Dimensions on page 7 Overall Dimensions 5 65 x 5 25 x 4 86 W x H x D 144 mm x 133 mm x 124 mm Weight 5 00 lbs 80 oz 2 27 kg Warranty 3 years parts and labor See Warranty Information and Terms Conditions on www predig com for complete details Input Input 4 20 mA Accuracy 0 03 of ...

Страница 6: ...oof for use in Zone 1 Ex d IIC T6 Gb Protection by Enclosure for use in Dust Atmospheres Zone 21 Ex tb IIIC T85 C Db IP68 Ta 40 C to 75 C Certificate number IECEx SIR 10 0056X Special Conditions for Safe Use Use suitably certified and dimensioned cable entry device and or plug The equipment shall be installed such that the supply cable is protected from mechani cal damage The cable shall not be su...

Страница 7: ...art is missing or the meter malfunctions please contact your supplier or the factory for assistance Pre Installed Conduit Plug The PD6820 is supplied with one pre installed conduit plug for installations that do not require the use of both conduit entries The conduit stopping plug includes an internal hexagonal socket recess for removal The conduit plug and its factory installation are included in...

Страница 8: ... mA signal return negative terminal connection when not using loop powered backlight BACKLIGHT 9 36 VDC when powering backlight from external supply BACKLIGHT 4 20 mA signal return negative terminal when using the installed loop powered backlight or ground negative when power ing backlight from external supply OUTPUT NPN open collector output positive OUTPUT NPN open collector output negative RESE...

Страница 9: ...xternal reset connections are made to two terminals labeled RESET Connect to a contact closure source such as a relay or a pushbutton as shown in Figure 7 Figure 7 Reset Connections Open Collector Output Connections Output connections are made to two terminals labeled OUTPUT Connect to an input device such as alarm indicator or pulse counter as shown in Figure 8 or drive a relay as shown in Figure...

Страница 10: ...re disabled when a mechanical button is pressed and will automatically be re enabled after 60 seconds of inactivity The SafeTouch buttons are designed to filter normal levels of ambient interference and to protect against false triggering however it is recommended that the SafeTouch buttons be disabled slide switch to LOCK if there is an infrared interference source in line of sight to the display...

Страница 11: ...cess the Advanced features of the meter Right Reset Button Press the Right arrow button to move to the next digit or decimal position during programming Up Display Button Press the Up arrow button to scroll through the menus decimal point or to increment the value of a digit Enter Button Press the Enter button to access a menu or to accept a setting The meter displays various functions and message...

Страница 12: ...ogram menu sCalE Scale Enter the Scale menu Cal Calibrate Enter the Calibrate menu Inpt1 Input 1 Calibrate input 1 signal or program input 1 value DspL1 Display 1 Program display 1 value Inpt2 Input 2 Calibrate input 2 signal or program input 2 value DsPl2 Display 2 Program display 2 value Span Error Span Error Error calibration not successful check signal TbASE Time Base Enter the Time Base menu ...

Страница 13: ... digit being changed blinks Press the Enter button at any time to accept a set ting or MENU button to exit without saving changes The decimal point is set using the Right or Up arrow button in the Setup decimal point menu Setting the Decimal Point dec PT Rate decimal point may be set with up to four decimal places or with no decimal point at all Total decimal point may be set with up to six decima...

Страница 14: ...te functions are exclusive of each other The meter uses the last function pro grammed Only one of these methods can be em ployed at a time The Scale and Calibrate functions can use up to 32 points default is 2 The number of points should be set in the Advanced Features menu under the Multi Point Linearization lnear menu se lection prior to scaling and calibration of the meter see Advanced Features...

Страница 15: ...e Enter key before calibration will take effect Minimum Input Span The minimum input span is the minimum difference between input 1 and input 2 signals required to complete the calibration or scaling of the meter The minimum span is 0 10 mA If the minimum span is not maintained the meter reverts to input 2 allowing the appropriate input signals to be applied Re Calibrating the Internal Calibration...

Страница 16: ...ess the Enter button at any time to accept a set ting or Menu button to exit without saving changes Setting the Tag Display tag The meter can be set to display a combination of sev en alphanumeric characters for engineering units e g GALLONS or for identification e g TANK 3 Press Right arrow to select next unit and Up arrow to increment unit To automatically cycle the lower display between total r...

Страница 17: ... to its password protect ed condition Disabling Password Protection To disable the password protection access the Password menu and enter the correct password twice as shown below The meter is now unprotected until a new password is entered If the correct six digit password is entered the meter displays the message UNLOCKD unlocked and the protection is disabled until a new password is programmed ...

Страница 18: ...t to rate Total Total Assign alarm output to total Set Set Point Program set point Reset Reset Point Program reset point Pulse Pulse Output Program pulse output Scaler Count MAX HZ Frequency Program pulse output maximum frequency Funct Function Enter advanced function menu Lnear Linear Set linear scaling Squar Square Root Set square root extraction Prog E Programmable Exponent Set programmable exp...

Страница 19: ... are programmed the same output will reset one count below set point Total alarm trip point program total set point Alarm reset is triggered by total reset There is no reset parameter entered for total If automatic total reset is enabled this setting will be the trigger point for the timer It is not necessary to have the output wired for automatic reset function to work To acknowledge an alarm pre...

Страница 20: ...a current source instead of using input scaling This process will override previously programmed scaling points Apply a live signal using a known accurate signal source InP 1 32 and then enter the corresponding meter reading disp 1 32 for that input signal level The use of a calibrated signal source is strongly recommended Important Navigation Note After entering the last display value the lineari...

Страница 21: ...eraging time and so the longer the display may take to find its final value The filter contains a noise filter bypass feature so that while small variations in the signal will be filtered out large abrupt changes to the input signal are dis played immediately Internal Calibration ICAL There is no need to recalibrate the meter for milliamps when first received from the factory The meter is factory ...

Страница 22: ...d hold Reset button when INFO is shown Press Enter when RESET DFALTS prompt is shown Note If Enter is not pressed within three seconds the prompt will stop flashing return to showing INFO Factory Defaults User Settings The following table shows the factory setting for most of the programmable parameters on the meter Next to the factory setting the user may record the new setting for the particular...

Страница 23: ... to exit Max Min display readings Display response is too slow Check filter setting to see if it can be lowered to LO or OFF If the display locks up or the meter does not respond at all Perform hard reset by shorting S and S terminals Backlight does not appear Backlight is intended for viewing assistance in dim lighting conditions It may not be noticeable under good lighting conditions Check conne...

Страница 24: ...ction or increment digit Cycle through maximum minimum and mA display mode ENTER Accept selection value and move to next selection Acknowledge alarms Max Min Mode While in Run Mode pressing Up Arrow will initiate MAX MIN Mode Up Arrow toggles between MAX MIN displays and Right Arrow resets the MAX MIN to the current value Press Menu or wait 10 seconds to return to Run Mode Pressing Enter Ack will ...

Страница 25: ...0079 0 2009 EN 60079 1 2007 EN 60079 31 2009 EN 61000 6 4 2004 EN 61010 1 2001 and EN 61326 2006 are no longer harmonized The requirements of these standards have been checked against the harmonized standard EN 55022 2010 EN 60079 0 2018 EN 60079 1 2014 EN 60079 31 2014 EN 61000 6 4 2019 EN 61010 1 2010 A1 2019 and EN 61326 2013 and there were no major technical changes affecting the latest techni...

Страница 26: ...5 8990 www predig com LIM6820_J SFT049 Ver 1 200 up 05 20 Contact Precision Digital Technical Support Call 800 610 5239 or 508 655 7300 Fax 508 655 8990 Email support predig com Sales Support Call 800 343 1001 or 508 655 7300 Fax 508 655 8990 Email sales predig com Place Orders Email orders predig com For the latest version of this manual please visit www predig com ...

Отзывы: