10
On the following screen you can change any protocol parameters:
Set the incubation time for blocking, primary (PA) and secondary (SA) incubation by touching adjacent white
area.
Note: The minimum time for blocking is 5 minutes, the minimum time for PA and SA incubation is 10
minutes.
Note: press
to skip the step, zero time for blocking, PA or SA will appear.
Set the number of washing cycles:
By touching the white area on left side of the screen set the number of washing cycles after blocking,
PA(primary Ab) incubation time and SA(secondary Ab) incubation time.
Note: you can skip washing after blocking by selecting zero, but you cannot skip washing after primary and
secondary antibodies incubation.
Set the duration of washing cycle:
By touching the white area on left of
“washing time”
you select the duration of each washing cycle. It
can be set between 3 and 20 min.
After finishing the protocol modification select side you would like to run:
.
Please note if the button is dimmed, this side is running and cannot be selected
Press
button on the top of the screen and use the following screen to verify protocol selection and to
start protocol (see below).