PRC-Saltillo AeroSwitch Скачать руководство пользователя страница 12

AeroSwitch User’s Guide 

 

12 

Explanation of Symbols 

This section explains the symbols that appear on the AeroSwitch or packaging. 

Symbol 

Explanation 

 

The switch conforms to European Union health, safety, and environmental 

standards. 

 

The entire switch is a type BF applied part. “Applied part” refers to the part 

of the switch with which the user comes into physical contact when using 

it for its intended function

.

 

 

Attention! Read all warnings and precautions in the instruction manual. 

IP44 

The switch is protected against solid objects over 1mm and water splashed 

in all directions. 

 

The range of temperatures to which the switch can be exposed while in 

use is 5°C to 40°C or 41°F to 104°F. During transport or while in storage, 

the range of temperatures to which the switch can be exposed is -25°C to 

70°C or -13°F to 158°F. See also “Environmental Conditions for Operating, 

Transport, and Storage” below. 

Caution:

 If the switch has been stored in an environment in which the 

temperature range is different from “the range of temperatures to which 

the device can be exposed while in use”, let the switch stand for at least 30 

minutes before using it. 

 

The range of humidity to which the switch can be exposed while in use is 

15% to 90%. During transport or while in storage, the range of humidity to 

which the switch can be exposed is 10% to 90%. 

 

Atmospheric pressure while in use is 70 kPa to 106 kPa. Atmospheric 

pressure while in storage is 50 kPa to 106 kPa. 

 

Consult instructions for use. 

 

The switch employs Bluetooth wireless technology. 

 

The date of manufacture of the switch and the name and address of the 

manufacturer. 

 

The authorized representative in the European Community. 

 

Specifications 

 

Dimensions 

2.99 in. (76 mm) in diameter and 0.8 in. (20 mm) in height 

Battery life 

Varies by user, switch mode setting, and how much the AeroSwitch is used 

 

Environmental Conditions for Operating, Transport, and Storage 

 

Operating 

Transport and Storage 

Temperature 

5°C to 40°C or 41°F to 104°F 

-25°C to 70°C or -13°F to 158°F 

Humidity 

15% to 90% 

10% to 90% 

Atmospheric Pressure 

70 kPa to 106 kPa 

50 kPa to 106 kPa 

 

Содержание AeroSwitch

Страница 1: ...10118S Version 1 00 AeroSwitch User s Guide ...

Страница 2: ...hitegates 25 27 High Street Swinstead Lincolnshire NG33 4PA UK Tel 44 0 1733 370 470 Fax 44 0 1476 552 473 Email info liberator co uk Web www liberator co uk Technical assistance Call Saltillo technical support at 1 800 382 8622 option 1 Call PRC technical support at 1 800 262 1990 Additional Information To view the AeroSwitch User s Guide videos FAQs troubleshooting tips warranty information and ...

Страница 3: ...ver provided with the switch could be a hazard for poking an eye Store it in a safe location when not in use Implantable Medical Devices A minimum of six 6 inches should be maintained between a handheld wireless mobile device and an implantable medical device such as a pacemaker implantable cardioverter defibrillator vagus nerve stimulator shunt or stent to avoid potential magnetic interference wi...

Страница 4: ...25 1 very short flash Less than 10 Consider getting a replacement battery Deciding on Broadcast Mode or Keyboard Mode The AeroSwitch has two connection options Broadcast mode and keyboard mode The switch ships in broadcast mode Broadcast Mode Keyboard Mode Limited to scanning WITHIN the communication software NuVoice Empower Chat software Able to run scanning both WITHIN and OUTSIDE the communicat...

Страница 5: ... NovaChat 10 6th generation NovaChat 12 Other devices with Bluetooth 4 1 and higher Setting up an Accent Device in Broadcast Mode 1 Ensure the AeroSwitch is turned off 2 Set the rotary dial to broadcast mode 0 3 Turn the AeroSwitch on 4 If your Accent device is running NuVoice software Go to the Toolbox and select Access Method Menu If switch scanning is not already selected select Choose Differen...

Страница 6: ...t One Switch or Two Switch 5 Select Switch 1 The prompt Press any switch will appear Press the preferred switch 6 If you selected Two Switch select Switch 2 and press the second switch Note To set up switch scanning refer to the NovaChat User s Guide or ChatFusion User s Guide provided with your device or available from https saltillo com support downloads Setting up a NovaChator ChatFusion Device...

Страница 7: ...scribed under the appropriate Setting up section above Disconnecting and or Unpairing the AeroSwitch and Device If you want to use the AeroSwitch with a different device unpair it from the original device Accent Device with NuVoice in Broadcast Mode 1 In NuVoice go to the Toolbox and select Access Method Menu 2 Select Disconnect ALL AeroSwitch 3 Select OK twice Accent Device with Empower in Broadc...

Страница 8: ...Switch is used Replacement batteries are available in stores Look for a CR2032 lithium coin cell battery Rechargeable options are available 1 Turn the AeroSwitch off 2 Remove the battery cover screws and cover Use the screwdriver provided with the switch 3 Remove the battery 4 Insert the replacement battery with the positive facing up toward you 5 Turn the AeroSwitch on If the blue LED flashes bri...

Страница 9: ...effectively disinfect it before another person uses it First Step Put on protective gloves Next Step Clean the AeroSwitch 1 Wipe down the switch with disinfectant wipes Virucidal Bactericidal Pseudomonacidal Tuberculocidal Fungicidal Metrex CaviWipes Follow the disinfectant product manufacturer s instructions for cleaning 2 Take a Q tip with cleaner applied for example Windex cleaner or another co...

Страница 10: ...hesive side firmly onto the switch mount 3 Fasten the switch to the mount making sure it is held firmly in place Attaching the AeroSwitch to a Switch Mounting Kit SMW 1 1 Line up the holes in the plate with the holes on the AeroSwitch 2 Insert the screws in the holes and tighten Other Mounting options Switch mounting kits are available from www prc saltillo com Finding Additional Information on Us...

Страница 11: ...Switch can be hand held operated on a table surface or mounted to a wheelchair The switch does not incorporate in any way medicinal substances human blood or plasma derivatives nor is it manufactured using tissues of animal origin Expected Service Life The expected service life of the switch is five years If you replace the battery in your switch dispose of the old battery in accordance with your ...

Страница 12: ...n environment in which the temperature range is different from the range of temperatures to which the device can be exposed while in use let the switch stand for at least 30 minutes before using it The range of humidity to which the switch can be exposed while in use is 15 to 90 During transport or while in storage the range of humidity to which the switch can be exposed is 10 to 90 Atmospheric pr...

Отзывы: