EZ-Read Micrometer

1. Functional Elements:

       (1) Locking device, (2) Digit counter, (3) Friction thimble, 

       (4) Sleeve, (5) Screw, (6) End cap, (7) Spanner

2. Zero Setting 

(If digital alignment is askew.)

:

       1.) 

Rotate friction thimble (3), set zero position for measuring face 

             just as normal micrometers.

       2.) Lock locking device (1), take off end cap (6).

       3.) Hold frame (for non-rotating spindle type micrometers, hold 

           sleeve (4)), loosen screw (5) with spanner (7) or a screw driver.

       4.) Rotate sleeve (4) until the reading on the digit counter (2) is OK. 

             The last rotating direction should be the same as measuring.

       5.) Hold frame (for non-rotating spindle type micrometers, hold 

           sleeve (4)), tighten screw.

       6.) Unlock locking device (1). Set zero position for measuring face 

           again. If the reading on digit counter (2) is not OK, repeat above 

           process.

       7.) Rotate friction thimble (3) backward to move the spindle back a 

           little. Hold friction thimble (3), tighten screw (5) firmly.

       8.) Replace endcap (6), make sure the end of the endcap (6) is in the 

keyway of the sleeve (4).

3. Precautions

       1.) Tighten screw (5) firmly to insure accuracy.

       2.) Do not disassemble and drop instrument.

       

1

2

3

4 5

6

7

!

CAUTION:

 

Excessive speed may cause internal gear damage!

EZ-Read Micrometer

1. Functional Elements:

(1) Locking device, (2) Digit counter, (3) Friction thimble, 

(4) Sleeve, (5) Screw, (6) End cap, (7) Spanner

2. Zero Setting 

(If digital alignment is askew.)

:

1.)  

Rotate friction thimble (3), set zero position for measuring face  

just as normal micrometers.

2.) Lock locking device (1), take off end cap (6).

3.) Hold frame (for non-rotating spindle type micrometers, hold  

sleeve (4)), loosen screw (5) with spanner (7) or a screw driver.

4.) Rotate sleeve (4) until the reading on the digit counter (2) is OK.  

The last rotating direction should be the same as measuring.

5.) Hold frame (for non-rotating spindle type micrometers, hold  

sleeve (4)), tighten screw.

6.) Unlock locking device (1). Set zero position for measuring face  

again. If the reading on digit counter (2) is not OK, repeat above  

process.

7.) Rotate friction thimble (3) backward to move the spindle back a  

little. Hold friction thimble (3), tighten screw (5) firmly.

8.) Replace endcap (6), make sure the end of the endcap (6) is in the  

keyway of the sleeve (4).

3. Precautions

1.) Tighten screw (5) firmly to insure accuracy.

2.) Do not disassemble and drop instrument.