Digital Micrometer
Part Number: 52-224 series
Safety Precautions
To ensure operator safety, use this instrument in conformance with the direc-
tions and specifications given in this User's Manual.
The edge of the contact point of this instrument is sharp.
Handle it with great care to avoid injury.
IMPORTANT
• Do not disassemble this instrument.
• Do not use and store this instrument at sites where the temperature will
change abruptly. Prior to use, thermally stabilize the instrument sufficiently
at a room temperature.
• Avoid storing this instrument in places where there is plenty of moisture
and dust, and using it at sites where it is directly subject to water or oil.
• Do not apply excessive force to this instrument or drop it.
• Be sure to perform the zero point adjustment before measurement.
• After use, wipe off dust, cutting chips, and moisture from this instrument, then
apply rust-preventive oil to it.
[1] Name and Function of Each Part
See the figure on the left.
[2] How to Read
See the figure on the left.
[3] Zero Point Adjustment
IMPORTANT
• Be sure to use the dedicated gauge block for the zero point adjustment (all of
these should be inspected periodically), the standard bar for micrometer, or
dedicated master gages for the workpiece to perform the zero point adjust-
ment of this instrument.
• Wipe off the measuring faces of the gage used and this instrument cleanly
before performing the zero point adjustment.
• Perform the zero point adjustment in the same posture and conditions as the
measurement.
1) When the reading is correct on the counter and wrong on
the thimble
(1) If the deviation is less than 0.02mm:
Rotate the sleeve with the supplied wrench by the observed error as in
the case for conventional micrometers. (Fig. 1)
(2) If the deviation is over 0.02mm:
Remove the ratchet stop with the supplied wrench. Push the thimble
outward to disengage it from the spindle, then rotate it to zero position.
(Fig. 2)
CAUTION
[1] Name and Function of Each Part
[2] How to Read
[3] Zero Point Adjustment
1. Frame
2. Spindle
3. Anvil
4. Sleeve
5. Thimble
6. Ratchet stop
7. Clamp
8. Counter
9. Wrench
Counter
Thimble
.0001" (Vernier)
.001"
.01"
.1"
1"
0-1" .0001"
0-1" .0001"
in.
Fred V. Fowler Co., Inc. • 66 Rowe Street • Newton, MA 02466
617-332-7004 • 617-332-4137 (fax) • Internet: www.fvfowler.com
Digital Micrometer
Part Number: 52-224 series
Safety Precautions
To ensure operator safety, use this instrument in conformance with the direc-
tions and specifications given in this User's Manual.
The edge of the contact point of this instrument is sharp.
Handle it with great care to avoid injury.
IMPORTANT
• Do not disassemble this instrument.
• Do not use and store this instrument at sites where the temperature will
change abruptly. Prior to use, thermally stabilize the instrument sufficiently
at a room temperature.
• Avoid storing this instrument in places where there is plenty of moisture
and dust, and using it at sites where it is directly subject to water or oil.
• Do not apply excessive force to this instrument or drop it.
• Be sure to perform the zero point adjustment before measurement.
• After use, wipe off dust, cutting chips, and moisture from this instrument, then
apply rust-preventive oil to it.
[1] Name and Function of Each Part
See the figure on the left.
[2] How to Read
See the figure on the left.
[3] Zero Point Adjustment
IMPORTANT
• Be sure to use the dedicated gauge block for the zero point adjustment (all of
these should be inspected periodically), the standard bar for micrometer, or
dedicated master gages for the workpiece to perform the zero point adjust-
ment of this instrument.
• Wipe off the measuring faces of the gage used and this instrument cleanly
before performing the zero point adjustment.
• Perform the zero point adjustment in the same posture and conditions as the
measurement.
1) When the reading is correct on the counter and wrong on
the thimble
(1) If the deviation is less than 0.02mm:
Rotate the sleeve with the supplied wrench by the observed error as in
the case for conventional micrometers. (Fig. 1)
(2) If the deviation is over 0.02mm:
Remove the ratchet stop with the supplied wrench. Push the thimble
outward to disengage it from the spindle, then rotate it to zero position.
(Fig. 2)
CAUTION
[1] Name and Function of Each Part
[2] How to Read
[3] Zero Point Adjustment
1. Frame
2. Spindle
3. Anvil
4. Sleeve
5. Thimble
6. Ratchet stop
7. Clamp
8. Counter
9. Wrench
Counter
Thimble
.0001" (Vernier)
.001"
.01"
.1"
1"
0-1" .0001"
0-1" .0001"
in.
7/6/04, 2:49 PM