Warning Notice
EGM-4 Operator's Manual – Version 4.19
- 6 -
Warning Notice
This instrument must not be used in situations where its failure could result in injury or
death. For applications where failure of this instrument to function correctly would
lead to consequential damage, the Analyzer must be checked for correct operation and
calibration at intervals appropriate to the criticality of the situation.
This manual is provided to help you install and operate the equipment. Every effort has been
made to ensure that the information contained in this manual is accurate and complete. PP
Systems does not accept any liability for losses or damages resulting from the use of this
information. PP Systems' equipment warranty is limited to replacement of defective
components, and does not cover injury to persons or property or other consequential
damage. This manual, and the information contained in it, is copyright to PP Systems. No
part of the manual may be copied, stored, transmitted or reproduced in any way or by any
means including, but not limited to, photocopying, photography, magnetic or other mechanical
or electrical means, without the prior written consent of PP Systems.
The EGM-4 is covered under warranty for one complete year, parts and labour included.
This, of course, is provided that the equipment is properly installed, operated and maintained
in accordance with written instructions (i.e. Operator's Manual).
The warranty excludes all
defects in equipment caused by incorrect installation, operation or maintenance,
misuse, alteration, and/or accident.
If for some reason, a fault is covered under warranty, it is the responsibility of the customer to
return the goods to PP Systems or an authorised agent for repair or replacement of the
defective part(s).
Condensation Risk
If condensation occurs in the analysis sample line, free water may enter the analyzer causing
serious damage which will result in a very expensive repair. Therefore, it is critical that you
periodically inspect the pipeline to be sure that condensed water does not enter the analyzer.
If the EGM-4 is supplied on its own, we do NOT supply any external filters.
If the EGM-4
is used in conditions where there is risk of condensation or water in the sample line, (i.e.
when used with the SRC-1, CPY-2, CPY-4 or CPY-3 chambers for measurement of
soil/canopy assimilation, we strongly recommend fitting our hydrophobic filter to the “Gas In”
port as shown below. Our hydrophobic filter assembly (Part No. 10045-1) is supplied with all
SRC-1’s, CPY-2’s, CPY-4’s and CPY-3’s. All EGM-4 customers can purchase this filter from
PP Systems (Part No. 10045-1) at any time if required. In addition, when this filter is used, it
must be checked regularly and replaced as when dirty or wet. Otherwise, blockage can
occur.
Our in-line filter assembly
(Part No. 10045-1)