I-Class
65
5.2.2 Advanced Entry Calibration
Advanced Entry is the alternate calibration method for special-case media types. In the
procedure, sensor readings for the label and TOF values are taken using different sampling
algorithms. From this compiled list of values the best algorithm is selected and then used to
generate new readings for manual entry into memory.
Note: Use this method only when Standard Calibration has proven unsuccessful.
With the correct SENSOR TYPE selected (see Section 4.2.2), perform an Advanced Entry
Calibration as follows:
Step
Action
Displayed Message
Comment
1
Turn ON the printer.
UNCALIBRATED
Wait briefly, about six
seconds, for the printer
to initialize.
2
Press the
Key, and
then raise the printhead
assembly.
MENU MODE
MEDIA SETTINGS
You are in MENU MODE.
3
Press the
Key.
MEDIA SETTINGS
MEDIA TYPE
You are in MEDIA
SETTINGS.
4
Press the
Key then
scroll to SENSOR
CALIBRATION.
MEDIA SETTINGS
SENSOR CALIBRATION
Press the
Key to
cancel this procedure.
5
Press the
Key, and
then the
Key.
SENSOR CALIBRATION
ADVANCED ENTRY
Press the
Key to
cancel this procedure.
6
Press the
Key, and
then the
Key.
ADVANCED ENTRY
SENSOR GAIN
You are beginning the
procedure.