The Salinity Pen

 Plus

is now ready for use!

 

It is recommended to recalibrate the Salinity Pen  

 

Plus+

 before each use.

Calibration of the Salinity Pen Plus

+

Calibration of testing equipment is essential for accurate 

and consistent results, and your new Salinity Pen

 Plus

+

 is no 

exception. All instruments are tested before they leave the 

factory however, it is good practice to calibrate your Salinity 

Pen

 Plus

+

 before use. 

Calibration Procedure:

Step 1

: Turn on the Salinity Pen

 Plus

+

 by depressing the

POWER

 button.

Step 2

: Press the 

Mode

 button until the display reads 

PPT

The calibration procedure 

WILL NOT WORK

 if the pen is not 

in 

PPT

 mode. 

Step 3

: Immerse the tip of the pen in the calibration solu-

tion. 

NOTE: When not in use, the calibration solution should be 

stored at room temperature.

Step 4:

 Press and hold the 

CAL/HOLD

 button for 10 full sec-

onds until you see the CAL message display on the LCD.

Содержание Salinity Pen Plus+

Страница 1: ...User Guide...

Страница 2: ...een The SP can also be easily calibrated in seconds using the included Salinity Standard Solution 30ppt This device has been factory tested to provide trouble free performance and contains certified c...

Страница 3: ...ct your local distributor directly for all warranty service and returns Visit poseidonflow com retailers Customer Support Should you require customer support for your Salinity Pen Plus help is o ered...

Страница 4: ...2 Button Batteries Included MODE Button Navigate between operational modes Temperature Celsius Fahrenheit Salinity PPT and Specific Gravity SG CAL HOLD Button Allows measurements to be locked on the d...

Страница 5: ...tinum sensor Clean only using a damp cloth or sponge with a neutral detergent DO NOT use chemicals such as acetone or benzene to clean any part of the instrument Avoid storing the instrument in places...

Страница 6: ...er it is good practice to calibrate your Salinity Pen Plus before use Calibration Procedure Step 1 Turn on the Salinity Pen Plus by depressing the POWER button Step 2 Press the Mode button until the d...

Страница 7: ...7 Correct Proper Salinity Pen Plus Usage Notice Measurement Keep the sensor at least 0 5 1cm away from all edges of the container when measuring Incorrect...

Страница 8: ...r 3 seconds until the unit change is observed on the display b Press the MODE button to toggle between testing modes Salinity ppt Specific Gravity SG and tem perature Celsius Fahrenheit 3 Immerse the...

Страница 9: ...with fresh water ensure to fully dry the sensor using a clean dry cloth or sponge Always keep the sensor dry when not in use WARNING Air bubbles stuck to the sensor after it is immersed into the wate...

Страница 10: ...chase receipt POSEIDON FLOW TECHNOLOGIES DISCLAIMS ALL OTHER WARRANTIES WHETHER ORAL OR WRITTEN EXPRESS OR IMPLIED TO THE FULLEST EXTENT ALLOWED BY APPLICABLE LAW NO IMPLIED WARRANTY OF FITNESS FOR A...

Страница 11: ......

Отзывы: