6
The Salinity Pen
Plus
+
is now ready for use!
It is recommended to recalibrate the Salinity Pen
Plus+
before each use.
Calibration of the Salinity Pen Plus
+
Calibration of testing equipment is essential for accurate
and consistent results, and your new Salinity Pen
Plus
+
is no
exception. All instruments are tested before they leave the
factory however, it is good practice to calibrate your Salinity
Pen
Plus
+
before use.
Calibration Procedure:
Step 1
: Turn on the Salinity Pen
Plus
+
by depressing the
POWER
button.
Step 2
: Press the
Mode
button until the display reads
PPT
.
The calibration procedure
WILL NOT WORK
if the pen is not
in
PPT
mode.
Step 3
: Immerse the tip of the pen in the calibration solu-
tion.
NOTE: When not in use, the calibration solution should be
stored at room temperature.
Step 4:
Press and hold the
CAL/HOLD
button for 10 full sec-
onds until you see the CAL message display on the LCD.