8. 

Install Front Plate with the six 4-40 flat head 
screws. 

9. 

Apply diameter indexed N

2

O/O

2

 gas label in 

place over diameter indexed check valve 
connector bushings.  Pull off center strip backing, 
and press label in place.  Remove top and 
bottom backing strips, and finish securing label.   

 

10. 

Connect an 8060 series duplex connector and 
hoses into the completely assembled Triple 
Outlet Station.  Attach the other end of the hoses 
to a flowmeter and turn the flow control knobs to 
the off position and the on/off switch to the off 
position. 

11. 

Leak test

 

the entire system for working 

pressure leaks

.

  Pressurize the sedation gas 

supply lines with 50 PSI.  Observe any pressure 
decay overnight. This 50 PSI test with the 
flowmeter tubing connections in place tests the 
seal of the duplex connector extended into the o-
rings of the outlet station primary check valves.  
(5 PSI drop allowed.) 

12. 

Attach the appropriate vacuum lines from the 
rubber goods to the outlet station vacuum source 
via the quick connect (5602).   

 

CONNECTIONS TO FLOWMETER – 
OPERATING INSTRUCTIONS 

Flowmeter Gas Supply Tubing and Vacuum Tubing is 
connected to the Triple Outlet Station via two quick 
connect couplers.  The Porter 8060 series duplex 
connector simultaneously connects the Oxygen and 
Nitrous oxide tubing, and prevents the Nitrous Oxide 
from being connected unless the Oxygen is 
connected.  The Porter 5602 vacuum quick connect 
with hose barb separately inserts into the vacuum 
check valve of the station.   

 
QUICK CONNECTING 

You may quick connect to the station when the 
system pressure is at its normal 50 PSI; the primary 
check valves seal this pressure.  Both couplers have 
locating latches.  Simply insert the couplers into the 
appropriate check valve and confirm the latch is in 
place behind the catch feature of the valve.  A gentle 
tug on the coupler will confirm a stable latch position.  
Internally, within the primary check valve, an o-ring 
seals against a poppet.  The same o-ring seals 
against the quick connect extensions upon insertion.   
Slide the latch out of position to remove the 
connector and the internal poppet will automatically 
move back into its sealing position. 

OXYGEN RESUSCITATOR 

By removing the duplex connector, an oxygen 
resuscitator quick connect may be inserted into the 
oxygen station position. 

 
 
MAINTENANCE AND SERVICE 

 

Use Scavenging 

Monitor for N2O in the operatory to insure that 
controls are effective in achieving low levels of ppm 
(parts per million) exposure.  Contact your Porter 
dealer for details on monitors and testing. 
 

Inspect and maintain

 the analgesia delivery 

system to prevent N2O leaks in all hoses, 
connections and fittings.  Repair all leaks 
immediately. 

 
 

Opposite position for 5602 right angle 
vacuum quick connect hose barb:

  

Reconfigure 

the fitting using the following steps if you need the 
barb to point in the opposite direction. 

Clip Reversal Instructions: 

1. 

Note orientation of Clip on Assembly:  Hold 

assembly with Hose Barb in the down position, 
and with the smooth post facing you. 

1.1.  Note the position of the clip on the 

assembly.  This is important, as you will 
need to rotate the clip 180

°

 from its 

original position. 

2. 

Remove the Crescent Ring using the blade of a 

small flat blade screwdriver. 

3. 

Lift the clip from the assembly, and as you do so, 

the spring will fall from the clip. 

4. 

Slide the clip back onto the assembly in a 

position 180

°

 from its original location. 

5. 

As the clip is brought down onto the post, place 

the spring onto the pin (located on the clip), and 
tuck the opposite end into the pocket on the 
adapter body.  

6. 

Again using a flat bladed screwdriver, push the 

Crescent Ring back into the groove on the post.  
Orient the Crescent Ring such that the opening in 
the ring faces away from the spring. 

7. 

Actuate clip to ensure proper operation of the 

assembly. 

 