User manual
PONSEL
Reference: NOTICE_ACTEON_2010_v001-UK
V: 001
42/56
9.3.2 Calibrating the 4-20mA outputs
Calibrating the 4-20mA outputs enables their precision to be improved by taking into account the load of your
device.
Use the
and
keys to select
the desired configuration.
Select the
4-20mA output calibration
menu then
press the ENTER key.
CONFIG
4-20mA OUTPUTS
4-20mA output adjustment
4-20mA output calibration
CONFIG
Procedure:
Insert an ammeter on 4-20mA
outputs
Once the 4-20mA output is correctly configured for the
pH, press the ENTER key.
Use the
and
keys to increase
and decrease the 4mA value.
Use the
and
keys to select
either the 4-20mA output for the pH
or the 4-20mA output for the
temperature.
CONFIG
4-20mA OUTPUTS
CURRENT GENERATION OF 4 mA
Enter the value
read on the
output 4-20mA pH
4.00
mA
Enter the value
read on the output
4-20mA temperature
4.00
mA
Esc: Escape the menu
Enter: To continue
Esc: Escape the menu
Enter: To continue