Parameter: Salinity (Please refer to above fig. C)
---
Meter is in manual
ranging 1 however the
conductivity measured
value is higher than
10.0ppt
Press “UP” key
more than 2
seconds to change
the mode to manual
ranging 2 or auto
ranging.
E03
Salt value is over the
range limit (70.0ppt) or
meter is damage
Put the meter in
standard buffer (the
buffer concentration
must be lower than
range limit). If E03
still appear, send
back for repair
E04
The original temp. or
conductivity error results
in this error
Refer to error code
of temp. &
conductivity. After
solving the error of
temp. &
conductivity, E04 of
TDS will disappear.
Parameter: Temperature (Please refer to above fig. D)
E01
Temperature circuit is
damaged
send back for repair
E02
Temp value is lower than
range limit (0
℃
) or
temperature circuit is
damaged
Put the meter in
room temp. for 5
mins to make the
readying back to
normal. If E02 still
appears, send back
for repair.
E03
Temp value is higher
than range limit (50
℃
) or
temperature circuit is
damaged
Put the meter in
room temp. for 5
mins. If E02 still
appears, send back
for repair.
23