NF802 Intelligent Relay Tester User Manual
20
4.2.1 SMV
It can make samples output and low level analog output in the format of collector in serial
based on IEC IEC61850-9-1, IEC61850-9-2, and IEC60044-7/8.
IEC61850-9-1
Press “IEC61850-9-1” in “SMV” to enter into the setting interface showing in figure 4.1.2.
Figure 0.2 IEC61850-9-1
Users are required to set the following parameters.
Sampling rate
: It is the same with the test object. Generally it is 200.
No. of ASDU
: It is the same with test object’s setting.
MAC (Dst. Address)
: It is the same with test object’s setting. Or it is set as broadcast
method with default value FFFFFFFFFFFF.
MAC (Sou. Address)
:Not set.
SCP
: Secondary current transformer for protection. Generally the default value is 463.
SCM
: Secondary current transformer for measurement. Generally the default value is