Problem
Cause
Corrective Action
No display
No power to unit
Mains Switch in Off
position
Check that power cord is
plugged into an operating
electrical outlet
Check that Mains Switch on
the rear of the unit is in the On
position
No heating
Set point lower than bath
temperature
Bath temperature above
High Limit temperature
Liquid level in bath is
too low
Check set point temperature;
increase as required.
Verify that High Limit
temperature is higher than
current set point temperature;
adjust as required (see page 8).
Check that liquid level is at
least 2 inches (5.08 cm) above
bottom of tank; add fluid as
required.
Insufficient
heating
Improper line voltage
Recent change in set
point or heat load
Bath cover not in place
Check that line voltage meets
specifications.
Allow sufficient time for bath
temperature to stabilize when
changes in heat load or set
point are made.
Check that bath cover is in
place.
Inaccurate bath
temperature
Incorrect calibration
Adjust Calibration Offset as
required (see page 10).
Troubleshooting Chart
Cleaning the Bath
Thoroughly clean the bath before each use. Use only mild soap and
water when cleaning. Do not use steel wool as damage to the unit may
result. Non-steel scouring pads are acceptable.
The entire unit is housed in a tough, well-insulated powder-coated steel
casing that is corrosion and chemical resistant.