Technical Information 

Performance Specifications 

Peak Working Temperature: 

-34.4°C (-30°F) 

Working Surface Dimensions: 

14.5 x 9.4 inches (36.8 x 23.8 cm) 

Overall Dimensions (L x W x H): 

15.9 x 18.4 x 10.5 inches (40.3 x 46.7 x 26.7 cm) 

Shipping Weight: 

76 lbs (34.4 kg) 

Electrical Requirements: 

120V, 60Hz, 6.5A 
240V, 50Hz, 3.5A 

 

Specifications subject to change without notice. 

 

Equipment Disposal (WEEE Directive) 

          or           

 

 
This equipment is marked with the crossed out wheeled bin symbol to indicate it is covered by the Waste 
Electrical and Electronic Equipment (WEEE) Directive and is not to be disposed of as unsorted municipal 
waste. 

Any products marked with this symbol must be collected separately, according to the 

regulatory guidelines in your area. 

It is your responsibility to correctly dispose of this equipment at lifecycle-end by handing it over to an 
authorized facility for separate collection and recycling. It is also your responsibility to decontaminate the 
equipment in case of biological, chemical and/or radiological contamination, so as to protect the persons 
involved in the disposal and recycling of the equipment from health hazards. By doing so, you will help to 
conserve natural and environmental resources and you will ensure that your equipment is recycled in a 
manner that protects human health. 
Requirements for waste collection, reuse, recycling, and recovery programs vary by regulatory authority 
at your location. Contact your local responsible body (e.g., your laboratory manager) or authorized 
representative for information regarding applicable disposal regulations.

 

 

Replacement Parts  

Description Part 

Number 

Operator’s Manual 

110-825 

Air Filter 

305-014 

Cover 

510-526 

Содержание Histology Freeze Plate

Страница 1: ...Operator s Manual Histology Freeze Plate 110 825 06 17 11 ...

Страница 2: ...acking 3 Package Contents 3 Controls and Components 4 Setup 5 Location 5 Electrical Power 5 Operation 5 Routine Maintenance 6 Cleaning 6 Air Filter 6 Technical Information 7 Performance Specifications 7 Equipment Disposal WEEE Directive 7 Replacement Parts 7 Service and Technical Support 8 Warranty 8 ...

Страница 3: ... involved in the installation operation or maintenance of the Histology Freeze Plate read this manual thoroughly prior to working with the unit This symbol marks chapters and sections of this instruction manual that are particularly relevant to safety When attached to the unit this symbol draws attention to the relevant section of the instruction manual This symbol indicates that hazardous voltage...

Страница 4: ...rature before cleaning or performing maintenance Use the cover provided with the unit to prevent accidental contact while not in use WARNING It is the user s responsibility to properly decontaminate the unit in the event hazardous materials are spilled on exterior or interior surfaces Consult manufacturer if there is any doubt regarding the compatibility of decontamination or cleaning agents Unpac...

Страница 5: ...110 825 4 Controls and Components Freeze Plate Cover Working surface Power Switch Circuit Breaker Ventilation openings IEC electrical connection Air filter access on bottom of unit ...

Страница 6: ...tension cord is not recommended However if one is necessary it must be properly grounded and capable of handling the total wattage of the unit The extension cord must not cause more than a 10 drop in voltage to the Histology Freeze Plate Operation Place the Power Switch Circuit Breaker on the front of the unit to the ON position The working surface will begin cooling and achieve working temperatur...

Страница 7: ... air filter can be accessed from the underside of the unit To keep the refrigeration system operating at optimum cooling capacity the front and back air vents and reusable filter should be kept free of dust and dirt They should be checked on a scheduled basis and cleaned as required Remove the filter for inspection and cleaning as follows 1 Position the unit so that the front overhangs the counter...

Страница 8: ...a It is your responsibility to correctly dispose of this equipment at lifecycle end by handing it over to an authorized facility for separate collection and recycling It is also your responsibility to decontaminate the equipment in case of biological chemical and or radiological contamination so as to protect the persons involved in the disposal and recycling of the equipment from health hazards B...

Страница 9: ...ges pre paid in either its original packaging or packaging affording an equal degree of protection to the indicated service center To insure prompt handling the return authorization number must be placed on the outside of the package A detailed explanation of the defect should be enclosed with the item The warranty shall not apply if the defect or malfunction was caused by accident neglect unreaso...

Отзывы: